Mechanical Specifications; Electrical Specifications - Gendex GXDP-700 Installation Manual

3d dental x-ray system
Hide thumbs Also See for GXDP-700:
Table of Contents

Advertisement

9 GXDP-700 Pre-sales check list
Installation space:
Pan/3D:
Minimum installation space
Depth: 1700 mm
Width: 1500 mm
Height: 2410 mm at minimum
(max height adjustable from
2140 to 2410)
Cephalostat:
Minimum installation space
Depth: 1700 mm
Width: 2500 mm
Height: 2410 mm at minimum
(max height adjustable from
2140 to 2410)4
Fixing hardware
Required fixing hardware de-
pends on the wall and floor
material and it is not delivered
with the unit. The fixing hard-
ware, as well as wall and floor
materials must endure a pull-
out force of 5000 N. Wall ma-
terial should be suitable for fix-
ing the unit. If the wall is made
of weak material, you m y have
to use a reinforcing plate on
the rear side of the wall to hold
the fixing hardware.The per-
son installing the unit is re-
sponsible for fixing hardware.
Check that it is possible to drill
holes in the floor without dam-
age to any electrical or water
pipes etc.
Weight:
The fully assembled unit with
ceph weighs 245 kg / 540 lbs
(212 kg / 467 lbs without
cephalostat). The area of the
column base plate is 0.07 m².
Make sure that the floor where
the unit is to be installed can
support this weight. The floor
should be able to carry mini-
mum 1000 kg/m
Power network
100-240VAC,
(10A@230VAC,
15A@110VAC)
power supply, max. 0.2Ω line
impedance. Separate outlets
for the unit and Workstation
PC.
154

9.3 Mechanical Specifications

(see image 1&2)
End User/Dealer
End User/Dealer
2
.

9.4 Electrical Specifications

50/60Hz
dedicated
End User/Dealer
Dealer