Orthoralix ® panoramic, we are dedicated to providing products with the features and high quality you desire. As you enjoy using the new DenOptix QST system, you are taking advantage of our 50 years of manufacturing expertise in the dental industry.
Supplies and Replacement Parts To order supplies or replacement parts for your DenOptix QST, contact your local Gendex equipment dealer. If your dealer is unable to assist you, call the Gendex location nearest to you. Indications for Use The DenOptix QST is intended for use as a digital dental radiography system using X-ray...
Chapter Important Information It is important that all personnel who will operate the DenOptix QST System read and understand this manual before operating the device. All personnel should follow all warnings and cautions as outlined below, for their safety and the safety of others around them.
Page 10
Safety - Collateral Standard; Electromagnetic Compatibility Requirements and Tests, class A. Use only grounded electrical connections. Connect the DenOptix QST scanner to a grounded electrical outlet between 100-240 volts AC. Do not reuse the barrier envelopes. Once used within the oral cavity, the barrier envelopes are bio-contaminated and should NEVER be reused.
Caution Statements Reduce the exposure time on your intraoral X-ray unit. The DenOptix QST System is designed to produce high quality, diagnostic images at reduced radiation levels. To get the maximum benefit, we recommend that the exposure be reduced as outlined in page 4-1.
DenOptix QST System Labels The following labels are attached to the system IEC Label Type B: Protection against electrical shock (IEC 60601-1) Caution Consult written instructions in this Manual. Warning This X-Ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed.
This manual provides the essential information necessary for the installation, operation and routine care of the DenOptix QST System. The detailed instructions for the imaging software associated with the scanner can be found in the Imaging Software User Manual and Installation Guide.
Page 14
DenOptix QST TIFF Tagged image file format LP/mm Line pairs per millimeter PhotoStimulable Phosphor D.P.I. Dots per inch Central processing unit (your computer) International Electrotechnical Commission Power on Power off Green indicator, ready Yellow indicator, scanning Red indicator, error in scanning; refer to operators...
Pre-Installation Information and Recommendations Physical Description The DenOptix QST system has several configurations. Even if the basic hardware is the same, DenOptix QST is provided with different licenses for scanning different types of plates, according to the chosen model. See the following table for a complete list of components for each model.
Green — Ready to scan • Yellow Blinking – Scanning • Red — Error Scanner Rear View 1. DenOptix QST scanner power entry module with switch and fuses 2. USB 2.0 connector. A USB 2.0 cable (provided) connects the DenOptix QST scanner to the CPU’s USB 2.0 port.
DenOptix QST Scanner Label Cool Gray Background Gendex® Dental Systems DenOptix® QST Digital Imaging System 100-240 V ~, 50/60 Hz, 100 VA MAX 0413 188046 250 V T2AH Gendex Dental Systems WARNING: For continued fire 1910 North Penn Road protection, use only specified type Hatfield, PA 19440 USA and rating of fuse.
Front (Blue) 1. IP serial number for tracking purposes 2. Date of manufacture 3. Available from Gendex Dental Systems 4. Solid orientation dot 5. IP size indicator. Shows the equivalent film size 6. Orientation circle. Visibly shows on final radiograph for faster orientation...
DenOptix QST Panoramic and Cephalometric Imaging Plates Panoramic Cephalometric Front (Blue) Rear (Black) 1. Plastic mounting strips used to hold panoramic imaging plate onto the carousel 2. Active imaging area, white color 3. Patient left ( ) and right ( ) side indicators 4.
Pre-Installation Information and Recommendations Carousels Extraoral QEO/Ceph (QCH) Intraoral (QFMX) 1. Top of the carousel. For the EO/CH carousels, the knob will be positioned based upon whether or not a Pan or Ceph is being scanned. If a Pan or Ceph is loaded, turn the knob towards the mark indicating a Pan or Ceph.
DenOptix QST Disposable Barrier Envelopes Back (Clear) Front (Blue) 1. The front of the barrier envelope is blue. This is the side that should face the X-ray tube. 2. Barrier envelope adhesive strip. When this is removed, the envelope can be sealed and provides protection from cross-contamination.
Pre-Installation Information and Recommendations Electrical Requirements Voltage 100-240 V AC 50/60 Hz power supply auto-senses the input voltage. Power 110 watts maximum Power Cord Standard line cord provided (Medical Grade for 115V) Compliance to Standards Standard Title UL 60601-1 Medical Electrical Equipment, Part 1: General requirements for safety 21 CFR Chapter I Subchapter J Performance Standard for Light-Emitting Products...
We do not recommend that an untrained person try to install and configure the system. Site Selection The DenOptix QST scanner can be located almost anywhere in the dental office. The site you pick should have: • Subdued lighting conditions. You should have the ability to turn down/off the lights and block sunlight with blinds.
Pre-Installation Information and Recommendations Unpacking the Unit The DenOptix QST System is shipped in one box. Completely unpack the box and save it in a safe, dry location. You may need to repack the unit for shipping if you ever encounter a maintenance outage, or relocate your office.
Step 4: Connect the device end of the USB 2.0 cable to the USB 2.0 connector on the back of the DenOptix QST scanner. Connect the scanner to a grounded power outlet. Make sure the other end of the USB 2.0 cable is connected to the USB 2.0 port on the...
Instructions Preparing Your Current X-ray Equipment The DenOptix QST System produces X-ray images of high quality and low noise yet offers a dose reduction in intraoral surveys up to 80% of the dose required for D-speed intraoral film. More details on how this technology works can be found in appendix A. Almost any intraoral or panoramic X-ray unit can be used with the DenOptix QST system.
Panoramic/Cephalometric Equipment The DenOptix QST system requires no adjustments to the panoramic/ cephalometric exposure settings (kV, mA or time). The DenOptix QST system can be used with almost any existing panoramic/cephalometric unit, regardless of age or manufacturer.
For more detailed information, see appendix B. Infection Control Prior to use, the DenOptix QST I/O imaging plates must be placed in barrier envelopes to minimize the possibility of cross-contamination. With proper application and use of the barriers, it will not be necessary to routinely cold sterilize the imaging plates. After erasing the imaging plates, follow these directions: 1.
Taking an X-ray 3. Ensure the computer, monitor and DenOptix QST scanner are switched on and properly connected. The green LED indicates that the system is ready to scan. 4. Launch your imaging software.
Page 31
Repeat until all I/O imaging plates are mounted. 8. Open the carousel well cover on the DenOptix QST scanner and insert the carousel. Close the lid. You are now ready to scan your I/O imaging plates.
DenOptix QST Panoramic/Cephalometric Imaging 9. Place the erased panoramic or cephalo- metric imaging plate into the cassette with the white side pointed toward the tubehead. Insert the cassette into the panoramic or cephalometric unit and expose in the usual manner.
Page 33
Push until you hear a click. Panoramic shown. 13. Open the carousel well cover on the DenOptix QST scanner and insert the carousel. Close the cover. You are now ready to scan your panoramic or cephalo- metric imaging plate.
DenOptix QST Scanning the Imaging Plate 14. Once the scanner is loaded and the patient file is opened, click the scanner icon on the imaging software toolbar. The software will sense whether an FMX, EO or Ceph carousel is loaded.
Page 35
System Operating Instructions 17. The procedure for scanning a panoramic/ cephalo- metric plate is the same as for the intraorals. It is possible to select how many plates to scan, and what size. As with intraorals, you can choose between 150, 300, and 600 DPI.
You are now ready to take an X-ray on the next patient. Power System ON and OFF The DenOptix QST System is designed to be left on continuously. The laser is only operational while the unit is actually scanning. If you wish to turn the system off, please remove the carousel and use the following procedure.
Maintenance Procedures Chapter The DenOptix QST System is designed for many years of trouble-free operation. It is manufactured from the highest quality components to ensure excellent performance. Maintenance that can be performed by the operator is minimal. Cleaning the System If the scanner, imaging plates or carousel becomes contaminated, clean in accordance with the directions given on page 4-10,”Preparing for the next patient”.
Page 38
DenOptix QST 1. Remove the intraoral imaging plate holder from the carousel by removing the three screws along the vertical plastic retention piece. After removing the three screws, the intraoral plate holder will come free from the carousel. 2. Spray the soiled surfaces of...
Operator Maintenance Scanner Maintenance The only part of the DenOptix QST scanner that may require operator maintenance is the fuse. If you experience any other service-related problem, contact your local Gendex dealer. Follow these directions to change the fuse.
Page 40
DenOptix QST 5. Turn the DenOptix QST scanner off as per page 4-10. Unplug the power cord and USB 2.0 cable. WARNING To avoid possible electrical shock, ensure that the power cord is unplugged. 6. Insert a small, flat tipped...
4-10. Carousel Maintenance The carousel from the DenOptix QST scanner can be maintained by the operator. If the intraoral imaging plate holder on an intraoral, panoramic or cephalometric carousel becomes damaged, it can be replaced by the operator. Order a replacement holder from your local Gendex dealer.
DenOptix QST Troubleshooting Trouble Probable cause Corrective Action No power Scanner not plugged in Check power cord at outlet and at the connection on the No green light back of the scanner. Turn on the scanner, then turn on the computer and monitor.
Page 43
If the green power indicator initialize when the turned on light is off, switch on the software is opened DenOptix QST scanner per pg. 4-10. The lid on the carousel well Close the lid. is open The USB 2.0 cable between Reconnect the cable.
Page 44
Transfer imaging plates from their protective barriers to the DenOptix QST scanner as quickly as possible using the Dark Box (within one hour of exposure). 032-0265-EN Rev 1...
Page 45
Imaging plates fall out The imaging plates have not Practice loading erased into the DenOptix QST been loaded properly. See imaging plates in full scanner carousel well page 4-4,” Taking and x-ray”...
Disposal of Waste Materials and Inoperative Parts Disposal regulations vary from country to country. Therefore, it is difficult to give specific instructions on the disposal of DenOptix QST waste materials and inoperative parts. In general, we believe the following guidelines to be true.
Then, follow the guidelines outlined in Chapter 4. Shipment The DenOptix QST System is designed to be shipped in its original shipping container by normal commercial carriers. Ensure that the carousel has been removed from the scanner and that the carousel well is empty.
(2) years from date of purchase. Parts will be repaired or replaced at our option. The DenOptix QST Scanner and Carousels must be installed and operated in accordance with the Gendex written instructions furnished with the unit.
Technical Specifications Chapter DenOptix QST scanner Height 39.4 cm 15.5 in Width 49.3 cm 19.4 in Depth 27.4 cm 10.8 in Weight (empty) 16 kg 35 lbs Interface Cables USB 2.0 cables Voltage 100-240 VAC Frequency 50/60 Hz Power 110 VA max.
Appendix Technology The DenOptix QST System produces high quality X-ray images at very low doses as compared to intraoral X-ray film. The imaging plates in the DenOptix QST Systems are much more efficient at capturing X-ray energy than film. This means that the timer on your intraoral X-ray source can be turned down to a very low level without affecting the image quality.
DenOptix QST How Storage Phosphor Imaging Plates Work Each DenOptix QST Imaging Plate is made up of a very thin layer of tiny storage phosphor crystals that are bonded together and coated on a flexible sheet of plastic. These storage phosphor crystals have the ability to capture the energy of X-rays and store the pattern as a latent image (see diagram).
Lighting Conditions for Appendix Imaging Plates Without light, storage phosphor imaging would not work. The red light from the laser scanner excites the exposed imaging plates, causing them to emit blue light. After your computer interprets the blue light and changes it into a digital image, light is used to erase the residual information from the imaging plate.
DenOptix QST If you cannot darken your mounting area to 20 lux or less, contact your local Gendex representative for suggested mounting times. Lighting Conditions for Erasing Imaging Plates In order to completely erase imaging plates, 99.5% of the image information must be removed.
Practical Test 1. Completely erase five size 2 imaging plates by subjecting them to direct, intense light for about 20 minutes. 2. Scan the imaging plates. No image should appear on the monitor. If an image does appear, repeat step 1 with a more intense light source. 3.
Optional Printer Appendix Because the DenOptix QST system can be interfaced with different applications software, and printer technology is constantly evolving, we can only suggest a list of suitable technical requirements for printers. To interface with the PC, the printer must have a digital input and not a video (analog) input.
DenOptix QST Application Software Please refer to the instructions in the application software for special information about printers. 032-0265-EN Rev 1...
Guidance and manufacturer’s declaration – electromagnetic emissions The Gendex QST is intended for use in the electromagnetic environment specified below. The customer or the user of the Gendex QST should ensure that it is used in such an environment. Emissions test...
Page 62
Guidance and manufacturer’s declaration - electromagnetic immunity The Gendex QST is intended for use in the electromagnetic environment specified below. The customer or the user of the Gendex QST should ensure that it is used in such an environment. Immunity test...
Page 63
Power 3A/m 3A/m Power frequency frequency magnetic fields should be at levels (50/60 Hz) characteristic of a magnetic field typical location in a IEC 61000-4-8 typical commercial or hospital environment NOTE: is the a.c. mains voltage prior to application of the test level 032-0265-EN Rev 1...
Page 64
Guidance and manufacturer’s declaration – electromagnetic immunity The Gendex QST is intended for use in the electromagnetic environment specified below. The customer or the user of the Gendex QST should ensure that it is used in such an environment. Immunity Test...
Page 65
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Gendex QST is used exceeds the applicable RF compliance level above, the Gendex QST should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re¬orienting or relocating the Gendex...