Enabling N2O/Air/O2 Operation - Dräger Medical Narkomed 6400 Operator's Instruction Manual

Air-only mode for the anesthesia system
Hide thumbs Also See for Narkomed 6400:
Table of Contents

Advertisement

4
Air-Only Mode Operation
Enabling N
O/Air/O
2
To exit Air-Only Mode and enable N
1. Press [To Secondary Keys] on the main screen taskbar.
Several control buttons change to choose additional tasks.
2. Press [N2O/Air/O2] on the secondary taskbar. The following dialog box
is displayed:
NOTE: N
Figure 4-8. Enable N2O/Air/O2 Confirmation Dialog Box
to cancel the request, touch [Cancel].
to enable N
actions occur:
4-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Operation
2
O/Air/O
operation can also be enabled immediately by
2
2
touching the [Enable N2O/Air/O2] button in certain dialog boxes
that appear while in Air-Only Mode. See "Special Conditions in
Air-Only Mode" on page 4-3.
O/Air/O
operation, touch [Continue]. The following
2
2
the minimum oxygen flow is enabled
the label of the control button on the secondary taskbar changes to
Air-Only Mode
the message N2O/Air/O2 is displayed under the oxygen
concentration value in the oxygen parameter box
the oxygen low alarm limit is set to 30%
the oxygen high alarm limit is set to 100%
Air-Only Mode Operator's Manual
RETURN TO CD-ROM TABLE OF CONTENTS
O/Air/O
operation:
2
2
Part Number: 4118081-001 Rev: A