Air-Only Mode
for the Narkomed 6400 Anesthesia System
WARNING: For a full understanding of the performance characteristics of this product, the
user should carefully read this manual before operating.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Narkomed 6400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Dräger Medical Narkomed 6400

  • Page 1 Rev: A Date: 30 January 2004 © 2004 Draeger Medical, Inc. Air-Only Mode for the Narkomed 6400 Anesthesia System WARNING: For a full understanding of the performance characteristics of this product, the user should carefully read this manual before operating.
  • Page 2 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 3: Table Of Contents

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Contents Section 1. Introduction Operator's Responsibility for Patient Safety ..........1-2 Limitation of Liability .
  • Page 4 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 5 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Introduction This section introduces you to the Air-Only Mode Operator’s Manual and provides important safety information. Operator's Responsibility for Patient Safety ....... 1-2 Limitation of Liability ............1-2 Restriction ................
  • Page 6: Operator's Responsibility For Patient Safety

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Introduction Operator's Responsibility for Patient Safety Draeger Medical anesthesia products are designed to provide the greatest degree of patient safety that is practically and technologically feasible. The design of the equipment, the accompanying literature, and the labeling on the equipment take into consideration that the purchase and use of the equipment are restricted to trained professionals, and that certain...
  • Page 7: Restriction

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Introduction whatsoever, regardless of any fault attributable to Draeger Medical and regardless of the form of action (including, without limitation, breach of warranty, negligence, strict liability, or otherwise). Draeger Medical shall in no event be liable for any special, incidental, or consequential damages (including loss of profits) whether or not foreseeable and even if Draeger Medical has been...
  • Page 8: Purpose Of This Manual

    All users of the Air-Only Mode must read this manual completely before using the equipment. This manual is intended to be used in conjuction with the Narkomed 6400 Anesthesia System Operator’s Instruction Manual . In addition to this introductory section, this manual contains three other sections: Section 2 - “General Description”...
  • Page 9: General Warnings And Cautions

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Introduction General Warnings and Cautions The following list of warnings and cautions apply to general operation of the Air-Only Mode. Warnings and cautions about specific functions appear with those topics.
  • Page 10 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 11: Section 2. General Description

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS General Description This section describes Air-Only Mode Option features and general functionality. Air-Only Mode Overview ............ 2-2 Air-Only Mode Control Button ..........2-2 Air-Only Mode Status Message ......... 2-3 Air-Only Mode Operator’s Manual Part Number: 4118081-001 Rev: A...
  • Page 12: Air-Only Mode Overview

    WARNING: Minimum oxygen flow is disabled in Air-Only Mode. Air-Only Mode Control Button The Air-Only Mode button is located on the secondary taskbar at the bottom of the main Narkomed 6400 screen. This button allows the clinician to toggle between Air-Only Mode and normal N O/Air/O operating mode.
  • Page 13: Air-Only Mode Status Message

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS General Description Air-Only Mode Status Message If the Air-Only Mode option is installed, a message indicating the status of the option is displayed in the oxygen parameter box. When Air-Only Mode is enabled , the message Air-Only Mode is displayed below the oxygen concentration value (See Figure 2-3.) This indicates that minimum oxygen flow has been turned off, allowing the clinician to use air...
  • Page 14 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
  • Page 15: Section 3. Preuse Checkout

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Preuse Checkout This section contains a preuse checkout procedure for Air-Only Mode. Preuse Checkout for Air-Only Mode ........3-2 Air-Only Mode Operator’s Manual Part Number: 4118081-001 Rev: A...
  • Page 16: Preuse Checkout For Air-Only Mode

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Preuse Checkout Preuse Checkout for Air-Only Mode Use this procedure daily and between successive cases to verify proper operation of Air-Only Mode. This checkout procedure should be used as part of the comprehensive checkout procedure provided in the Narkomed 6400 Operator’s Instruction Manual .
  • Page 17 RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Air-Only Mode Operation This section contains operational information for Air-Only Mode. Enabling Air-Only Mode ............. 4-2 Special Conditions in Air-Only Mode ......... 4-3 Loss of O2 Numeric Data in Air-Only Mode ....4-3 Significant N2O/O2 Levels Detected in Air-Only Mode 4-5 New Case in Air-Only Mode ........
  • Page 18: Enabling Air-Only Mode

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Air-Only Mode Operation Enabling Air-Only Mode Air-Only Mode can be enabled only when there is valid oxygen data and when the oxygen sensor is properly calibrated. If the [Air-Only Mode] button is touched when there is no valid oxygen reading, the hint message “A valid oxygen concentration must be displayed to select Air-Only Mode”...
  • Page 19: Special Conditions In Air-Only Mode

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Air-Only Mode Operation to cancel the request, touch [Cancel] . • • to enable Air-Only Mode, adjust N O flow to zero, appropriately adjust oxygen flow, and touch [Continue]. The following actions occur: –...
  • Page 20 Figure 4-3. Loss of Numeric Data in Air-Only Mode (Calibration Problem) • Calibrate the oxygen sensor or replace the oxygen sensor capsule. Refer to the Narkomed 6400 Anesthesia System Operator’s Instruction Manual for specific instructions. • Touch [Continue] to remove this dialog box. The system remains in Air- Only Mode.
  • Page 21 (so that all three gases can be delivered), touch [Enable N2O/Air/O2] . Significant The following dialog box appears on the screen if the Narkomed 6400 detects a nitrous oxide concentration higher than 5% or an oxygen Levels concentration higher than 25% while operating in Air-Only Mode:...
  • Page 22: New Case In Air-Only Mode

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Air-Only Mode Operation New Case in The following dialog box appears on the screen if the [New Case] control Air-Only button is touched while in Air-Only Mode: Mode Figure 4-6.
  • Page 23: Vpo Hardware Fault In Air-Only Mode

    Air-Only This dialog indicates that a hardware fault has occurred and the Mode Narkomed 6400 system monitor is unable to determine whether the minimum oxygen flow delivery is truly off. WARNING: Oxygen concentration levels can rise if control of minimum oxygen flow delivery is lost.
  • Page 24: Enabling N2O/Air/O2 Operation

    RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Air-Only Mode Operation Enabling N O/Air/O Operation To exit Air-Only Mode and enable N O/Air/O operation: 1. Press [To Secondary Keys] on the main screen taskbar. Several control buttons change to choose additional tasks. 2.