18 Maintenance

All maintenance work must be performed by qualified personnel or a Service Technician.
WARNING
Infection due to contaminated unit
Clean and disinfect the suction system before working on the unit.
Wear protective equipment during your work (e. g. impermeable gloves, protective goggles
and mouth and nose protection).
Prior to working on the unit or in case of danger, disconnect it from the mains.
18.1

Mojave Mobile

Maintenance interval
Weekly
According to need
Annually
18.2

VS 300 S

Maintenance interval
Every 4 weeks
Annually
*
May be done by Service Technicians only
18.3

VSA 300 S

Maintenance interval
Every 4 weeks
Dependent upon the
level of usage of the
device
ATH7100005L29 2011V004
Maintenance work
❯ Change the yellow disposable filter in the hose manifold.
❯ Change suction hoses
❯ Grease O-rings
❯ Change exhaust air filter.
Maintenance work
❯ Check the protective strainer on the suction connection of the unit and
clean or replace it according to need.
❯ Check the inlet and outlet hoses for signs of deposits/blockage or cracks
and replace them where necessary. *
❯ Check the outflow valve and replace it if necessary. *
Maintenance work
❯ Check the protective strainer on the suction connection of the unit and
clean or replace it according to need.
❯ Change the amalgam collecting container when a fill level of 95% or
100% is indicated on the display panel
Notes concerning prophy powders:
The amalgam separator is not functionally affected by conventional prophy
powders. Under certain circumstances however, increased soiling of lines
and hoses and a more frequent need to change the amalgam collecting
container can be expected.
Usage