Cleaning Sip-N-Puff - Invacare LiNX DLX-REM500 User Manual

Table of Contents

Advertisement

Invacare® LiNX
Remove screw/hand screw A and backrest shroud B.
2.
Remove saliva trap C from tube.
3. Insert new saliva trap with INLET imprinting facing
towards input module.

6.3 Cleaning Sip-N-Puff

Risk of Damage to Input Module
Improper mounting or maintenance of Sip-N-Puff
control may cause damage to input module by
water or saliva.
– Mouthpiece and breath tube MUST be
completely dry before installation.
Cleaning at least twice a week is recommended.
116
Fig. 6-3
1.
Remove mouthpiece A and lipswitch C from gooseneck
B.
2. Remove breath tube from saliva trap. See 6.2 Replacing
Saliva Trap, page 115 .
3. Position catch can beneath breath tube to collect water
and rinse.
4. Flush mouthpiece and breath tube with warm running
water.
5. Rinse with oral rinse to disinfect.
6. Let dry completely before installation.
7. Install mouthpiece, lipswitch and breath tube.
Fig. 6-4
1637425-I

Advertisement

Table of Contents
loading

Table of Contents