Equipment Required; Preparation; Disconnection - ABB AquaMaster 3 FET200 User Manual

Electromagnetic flowmeter transmitter
Hide thumbs Also See for AquaMaster 3 FET200:
Table of Contents

Advertisement

AquaMaster 3 FET200
Electromagnetic flowmeter

4.4.2 Equipment Required

Cleaners are available from your local ABB representative. To purchase supplies directly or for local
distributor details please go to the following website:
http://store.caig.com/
Material details are:
Description
DeoxIT
®
– Contact Cleaner & Rejuvenator
DeoxIT
®
– Mini-spray, 5 % solution, flushing action, 14 g
(Applications = 150 approx.)
®
DeoxIT
GOLD – Contact Enhancer, conditioner & Protector
DeoxIT
®
GOLD G5 Mini Spray 5 % solution, 14 g, flushing action and safe on plastics
(Applications = 150 approx.)

4.4.3 Preparation

Item
Real-time Clock
Transmitters with Data
Loggers

4.4.4 Disconnection

Before DeoxIT treatment, disconnect ALL cables in the following order:
1. Battery pack / power
2. Sensor
3. Pressure transducer (if fitted)
4. Outputs
5. Communications cable (if connected)
Uncap unused connectors.
OI/FET200–EN Rev. N
Precaution
This procedure may result in the loss of the real-time clock.
Once the treatment is complete, check, and if necessary, re-program the
real-time clock and date – see Section 4.4.8, page 31.
This procedure may result in the loss of logger contents on transmitters fitted
with data loggers.
To prevent data loss, download logger data before treating the connector
pins.
Part No.
D5MS–15
G5MS–S
29