APPENDIX A – ELECTRICAL CALIBRATION
Note. The instrument is calibrated by the company prior to despatch and an electrical calibration should be carried out
if the instrument's accuracy is suspect
suitably calibrated test equipment is available.
only
and
A1 Equipment Required
a) Millivolt source (cell input simulator), –20.0 to 180.0mV.
b) Millivolt source (temperature input simulator), 10.0 to 40.0mV.
c) Digital voltmeter (current output), 0 to 20mA.
d) Thermometer, to measure ambient temperature.
A2 Preparation
a) Switch off the mains supply. Disconnect the probe and retransmission output terminations from the instrument – see Fig. 4.4
or 4.5.
b) Connect the millivolt sources and the milliammeter to the appropriate terminals – see Fig. 4.4 or 4.5.
c) Switch on the mains supply to the instrument and allow ten minutes for the circuits to stabilize.
d) Select the
and proceed as detailed in Section A3.
Electrical Calibration Page
22