Appendix B - EMC test levels.
Test standard
CISPR11:2009+A1:2010
CISPR11:2009+A1:2010
IEC/EN 61000-4-2
IEC/EN 61000-4-3
IEC/EN 61000-4-3
IEC/EN 61000-4-4
IEC/EN 61000-4-5
IEC/EN 61000-4-6
IEC/EN 61000-4-11
Description
Radiated emissions
Conducted
emissions
Immunity from
electrostatic
discharge
Radiated RF
immunity
Radiated immunity
from intentional
transmitters
Immunity from
electrical fast
transients and bursts
Surge immunity on
AC supply
Conducted RF
immunity
Immunity to voltage
dips, short
interruptions and
voltage variations
OM265EN issue 16
Class/Group/Immunity
test level
Class A Group 2
Class A Group 2
±15kV air, ±8kV contact
3V/m
28V/m maximum
±2kV AC supply line,
±1kV signal lines
±2kV common mode,
±1kV differential mode
3V rms 150kHz > 80MHz,
6V rms ISM and amateur
bands
10ms > 5s dip/interruption
time