24h Blood Pressure Monitor boso TM-2450

Error codes

Error code Cause
0:00
Time sets to 0:00 when battery is replaced
E03 / E90
Zero point adjustment not possible
E04
Empty battery
E05
Leakage
E06
Pressure above 299 mmHg
E07
User abort via START/STOP button
E08 / E10
No or non-interpretable oscillations
E09
Error of the activity sensor
E20
Pulse < 30 or > 200
E21
No interpretable oscillations
E22
In the area of diastole (E21) or systole (E22)
E23
Systole-diastole < 10 / > 150 mmHg
E30
Measuring time longer than 180 seconds
E31
Deflation longer than 90 seconds
E48
Pulse cannot be measured
E52
Storage error
E91
Pressure inside the cuff too high /
maximum pressure set too low
Remedy
Device must be reprogrammed.
Deflate cuff completely.
Charge or replace batteries.
Disconnect the cuff from the device and reconnect.
If the error occurs repeatedly, contact your sales partner.
The arm must be kept still during measurement.
The arm must be kept still during measurement.
Remove and re-insert the battery packs.
Check the position and fit of the cuff.
Check the position and fit of the cuff.
Check the position and fit of the cuff.
Contact your sales partner.
Contact your sales partner.
The arm must be kept still during measurement.
Contact your sales partner.
Set a higher maximum pressure. The arm must be kept still
during the measurement.