Error messages

If any problems occur during the measurement, the display will show an error message
instead of a blood pressure reading.
Meaning of error messages:
GB
54
Cause of Err message:
Excessive variation in cuff pressure Hold arm still
during measurement
No valid pulse readings
Systole – diastole difference
too small:
Systole – diastole < _ 10 mmHg
Cause of Err CuF message:
Incorrectly inflated
Troubleshooting:
Check position of cuff (see page 47)
and repeat measurement.
Check position of cuff (see page 47)
and repeat measurement.
Troubleshooting:
Cuff possibly too loose, attach
cuff more firmly (see page 48)