24h Blood Pressure Monitor boso TM-2450

Important note

To ensure proper functioning of the boso TM-2450 over 24 hours, only
use rechargeable batteries rated at least 1900 mAh, 1.2 V NiMH.
In addition to the two rechargeable batteries required for the power
supply, the boso TM-2450 also contains an internal battery to save the
measurement data and the program settings in the device.
To prevent any loss of programming and stored measured
values when the internal battery is discharged, lease follow
the following procedure:
- Insert charged batteries into the device even when not in use.
- The state of charge of the internal battery is thus constantly kept
at a high level. If the internal battery's power supply is interrupted,
the settings of the boso TM-2450 will be lost after approx. 10 days.
- Before attaching the device to a patient, please replace the batteries in the
device with a set of freshly charged batteries.