M&C PSS SS-5 Instruction Manual page 16

Gas conditioning unit
Table of Contents

Advertisement

All parts which require maintenance work are housed in the gas conditioning system in such a way so
that they are easily accessible. These are (see figure 2):
 The filter element of the preliminary filter FP-2T.
In order to protect the analysers which have been additionally connected, it
is recommended that in the event of a condensation irruption the moist filter
elements be replaced.
N O T E !
 The preliminary filter for the peristaltic pump PF2. If the condensate contains particle residue, the
preliminary filter should be replaced at regular intervals. The 'one-way' filter is situated in the suction
side of the pump hose (see figure 2) and can be easily replaced;
 Hoses of the Condensate pump SR25.1. These should be checked every six months and, if
necessary, replaced;
 Diaphragm of the sample gas pump N3KPE or N9KPE. These should be checked every six months
and, if necessary, replaced;
16
SS-5, SS-5/3 | 1.00.01
www.mc-techgroup.com