Connect the hose end of the instrument air or check gas bottle connection to the gas inlet
of the analyser.
Open the pressure reducer valve slowly, to avoid pressure peaks.
Adjust the flow rate to 50 Nl/hr at the flow meter.
N O T E !
Wait approximately 20 – 30 sec. until the indication has stabilised.
If necessary adjust span accurately according to the check gas concentration with a screw
driver at the span potentiometer in the front panel. In case of air e.g. to 20,9 % O
Check output signals at 20,9% O
Output signal
0-1 V
off
0-20 mA
4-20 mA
bottle- resp. Instrument air valve resp. integrated sample gas pump.
Disconnect hose connection at the analyser.
Determination of the output signal:
– S
(S
) V resp. mA x gas concentration Vol % O
e
np
upper range value Vol % O
Se = Final value, signal output
Snp = Zero, signal output
Shut off pressure reducer output valve and bottle valve. Druckreglerausgangsventil und Fla-
schenventil schließen. Disconnect hose connection at the analyser.
The span calibration is finished.
N O T E !
A T T E N T I O N
20
Always calibrate at the flow rate that is adjusted for the measurement
too.
:
2
2
If during the span calibration great variations have to be compensated
(>2% O
) at the potentiometers, a second zero and span calibration is
2
reasonable.
After finishing the calibration turn measuring selection switch to the de-
sired measuring range.
The mA-output signal is dependant on the measuring range!
PMA20 | 1.00.01
Measurement
range 100 % O
2
0,209 V
4,18 mA
7,34 mA
2
+ S
np
Measurement
range 30 % O
2
0,697 V
13,93 mA
15,15 mA
www.mc-techgroup.com
.
2
Shut
pressure
reducer
valve and