Calibration; Zero Calibration - M&C PMA Series Instruction Manual

Hide thumbs Also See for PMA Series:
Table of Contents

Advertisement

After approximately 30 min. the LED in the indication instrument starts flashing. After 12
hours the transmitter has reached a temperature balance and the analyser is ready for cal-
ibration.
16

CALIBRATION

The accuracy of an analyser mainly is dependent on its calibration.
Before calibration it has to be assured that the calibration conditions cor-
respond to the conditions during measurement. The flow rate and the bar-
ometric pressure conditions have to be constant. Under this terms a cali-
N O T E !
bration of the analyser is necessary aproximately one time a week to main-
tain the accuracy. If flow rate or barometric pressure conditions are chang-
ing significantly, a new calibration is necessary.
During calibration the device must not be exposed to vibrations.
For zero calibration of the analyser an oxygen-free gas, mostly nitrogen (N
For span calibration with M&C O
because of the measuring principle and the linear measuring ranges. Dry and clean air is sufficient.
For measurement concentrations > 40% O
possibly recommandable.
16.1

ZERO CALIBRATION

Connect a flexible PVC- or FEP-hose with the pressure reducer of the N
The pressure reducer should have an output control range of max. 0 – 1,5 bar abs..
C A U T I O N !
The outlet pressure is only allowed to be adjusted at max. 0,1 bar. Other-
wise the measuring cell of the analyser will be destroyed.
Open the bottle valve and than the closed pressure reducer outlet valve and purge the
pressure reducer and the complete hose line for approximately 5 sec. the pressure reducer
and the complete hose line.
Check the adjusted control pressure and reduce if necessary to ≤ 0,1bar, then shut off the
pressure reducer valve again.
Connect the hose end of the zero-gas bottle connection to the gas inlet of the analyser.
Open the pressure reducer valve slowly, to avoid pressure peaks.
Adjust the flow rate to 50 Nl/hr at the flow meter.
Always calibrate at the flow rate that is adjusted for the measurement too.
N O T E !
Wait approximately 20 – 30 sec. until the indication has stabilised.
Turn measuring range selection switch to position 3%
If necessary adjust zero accurately to 0 % O
in the front panel
www.mc-techgroup.com
-analysers it is possible to abandon on special test gas mixtures
2
a calibration with corresponding test gas could be
2
with a screw driver at the zero potentiometer
2
PMA20 | 1.00.01
5.0) is used.
2
-zero-gas bottle.
2
15