MEASURING

For the first starting up at a new location, all steps in chapter 16 and 17 have to be performed.
By the requirements of precision the interval of the new calibration can be carried out daily or
weekly.
The sample gas must be free from all liquid or solid particles, i.e. the dew
C A U T I O N !
point of the gas has to be below the equipment temperature so that no
condensate will occur inside the equipment. If necessary, lower the dew
point by means of a cooler or dryer. For dust filtration use a filter of 2
micron porosity !
We will be pleased to inform you about an optimal gas conditioning
The analyser now is ready for operation.
18

CLOSING DOWN

In case of a short time closing down of the the analyser no further precautions are required.
In case of a closing down of the analyser for a longer period, it is recommended to flush the ana-
lyser with dry and clean inert gas (eg. surrounding air) in order to prevent a damage of the meas-
uring cell by aggressive and corrosive liquid gases.
19

MAINTENANCE

The analyser, woking with a physical measuring principle, requires no intensive and complex
maintenance. But the preceding components necessary for the sample gas conditioning are to be
maintained with special attention according to the respective operating manuals.
The calibration of zero and span is to be effected with the corresponding test gases according to
stability of the operating conditions and to the demands on the accuracy. Recommended interval
of calibration for standard applications: 1 x per week.
20

TROUBLE SHOOTING

Error
Possible reason
No indication
No supply voltage
No sample gas
Sample line or filter is
flow
blocked
Contamination of the inter-
nal diaphragm pump
www.mc-techgroup.com
Check/Repair
Check supply voltage according to type plate.
Check wether mains cable is plugged in accurate.
Check fine fuse in the low heat device socket.
Check sample system.
PMA20 | 1.00.01
21