The Measuring Principle; Figure 4 Scheme Of The Measuring Cell And Optical Signal Processing - M&C PMA Series Instruction Manual

Hide thumbs Also See for PMA Series:
Table of Contents

Advertisement

10

THE MEASURING PRINCIPLE

Oxygen is a gas with a significant paramagnetic susceptibility. The molecules of oxygen are at-
tracted much more strongly by a magnetic field than the molecules of other gases.
The measuring principle shown in the following is benefitting from these characteristics of the ox-
ygen. The great advantage of the paramagnetic measuring principle is the highly reduced cross
sensitivity of the measurement to other components in the sample g as.
Figure 2 shows the diagram of the measuring cell as well as the optical system for the detection of
the dumbbell's movement.
1
5
Figure 4
The measuring cell consists of two nitrogen-filled spheres  which are arranged in the form of a
dumbbell. In the dumbbell's central point of rotation, a small mirror  is placed. The dumbbell is
surrounded by a wire coil needed for the compensation procedure. The described system is fixed
rotationally symmetrical inside a glass tube via a tightening strap out of platinum  and is srewed
up with two pole pieces .
Two permanent magnets are producing an inhomogeneous magnetic field. When oxygen is flowing
in, the molecues of the oxygen are drawn into the magnetic field. In consequence, the lines of
electric flux on the cuneiform pole pieces  are compressed. The nitrogen-filled diamagnetic
sheres are pushed out of the magnetic field. This causes a rotation of the dumbbell. The rotation
is detected via an optical system consisting of mirror , projection LED  and photoelectric cell
.
In case the dumbbell is pushed out of the magnetic fieild, the tension of the photoelectric cell is
immediately changed. The measuring amplifiers  and  are producing a respective current which
develops via the wire coil on the dumbbell an electro-magnetic load moment. The load moment is
resetting the dumbbell into its zero position.
www.mc-techgroup.com
4
2
3
6
Scheme of the measuring cell and optical signal processing
GAS INLET
GAS OUTLET
7
8
PMA20 | 1.00.01
9
11