Download Print this page
SYSMEX BEYOND A BETTER BOX RU-21 Quick Manual

SYSMEX BEYOND A BETTER BOX RU-21 Quick Manual

Reagent preparation unit

Advertisement

Quick Links

B E YO N D A B E T T ER B OX
RU-21
TM
REAGENT PREPARATION UNIT
QUICK GUIDE

Advertisement

loading

Summary of Contents for SYSMEX BEYOND A BETTER BOX RU-21

  • Page 1 B E YO N D A B E T T ER B OX ™ RU-21 REAGENT PREPARATION UNIT QUICK GUIDE...
  • Page 2 RO (Reverse Osmosis) Water Specifications The RU-21 requires Type II reagent grade water based on CLSI Reagent Water Specifications*. Water quality is verified by Sysmex pre-installation. After installation, it is the laboratory’s responsibility to check and maintain water quality on a regular basis.
  • Page 3: Reagent Replacement

    REAGENT REPLACEMENT Reagent Replacement “CELLSHEATH CCS” - With Error A Help window from the RU-21 menu will display “Replace CELLSHEATH CCS” in the error message list. 2. Click [OK] to display the Replace Reagent window. Click the remaining reagent volume indicator for CELLSHEATH CCS. 4.
  • Page 4 RU-21 — REAGENT REPLACEMENT Troubleshooting - Switch from CELLSHEATH CCS to CELLSHEATH (SE-90L) (For Emergency Use Only) The switch to CELLSHEATH may be made in the following circumstances • Problem with the RO water / RO water system • Problem with the RU-21 Connect CELLSHEATH via Supply Tank Locate the Supply Tank.
  • Page 5 REAGENT REPLACEMENT Troubleshooting - Extended RU-21 / Type II Water System Downtime If you have made the switch from CELLSHEATH CCS to CELLSHEATH (SE-90L) and that cube runs out, you will need to replace it. Reagent Replacement - CELLSHEATH to CELLSHEATH (SE-90L) (For Emergency Use Only) The ‘Help’...
  • Page 6 RU-21 — TROUBLESHOOTING Pressure Adjustment [0.07 MP a Pressure Error ] Click the RU-2 menu button in the control menu. 2. Click the submenu button. Click [Show Status]. • The current air pressure will be displayed in the Show Status window from the RU menu •...
  • Page 7: Troubleshooting

    TROUBLESHOOTING Troubleshooting at the RU- 21 If an error occurs in the RU-21 an error message will appear. An error that has occurred can be cleared from the error list, which is opened by touching the Help button. To clear the error, touch [OK] to close the Error message, then touch the [Help] button. 2.
  • Page 8 B E YO N D A B E T T ER B OX ™ Sysmex delivers a comprehensive and intuitive ecosystem that improves performance with next generation diagnostics, advanced tools and technologies, process optimization and BeyondCare harmonized support. Sysmex America, Inc.