Clarity Platinum Urine Analyzer
OPERATOR'S MANUAL
POTENTIAL SOURCES
OF ERROR
Use of incorrect
QC controls/
expired controls
or improperly
stored and
handled controls
Unauthorized
operator
We Support Better Outcomes™
VALIDATION RESULTS
Users will receive a 'Barcode Error' message if they
use the incorrect control level, or if they are using
an expired bottle of controls.
The system will not proceed to the Quality Control
test, and the user will need to use an unexpired
bottle of controls or the correct control level to get
to the quality control test function.
The user will be alerted with a 'QC Fail' error
message if they select an improperly stored control
that was compromised due to incorrect storage
conditions. The system is locked out and the user
cannot perform any patient testing until the Quality
Control testing is completed successfully.
The user will be alerted with an 'Operator Login
Invalid' error message if the username or password
is incorrect.
Page 25