Advanced Settings; Operator; Test Mode; Quality Control - Clarity Platinum Operator's Manual

Urine analyzer
Table of Contents

Advertisement

4.2 Advanced Settings

Press the Advanced
settings menu. From this menu, it is possible to perform quality control testing,
strip bed maintenance, and system diagnostics on analyzer hardware.

Operator

The operator menu is only accessible for Clarity Diagnostics administrators.

Test Mode

The test mode menu is only accessible for Clarity Diagnostics administrators.

Quality Control

On the Advanced settings menu, press the Quality Control
the quality control menu. From here, operators can perform Quality Control
testing.
The QC settings option is only accessible for Clarity Diagnostics administrators.
The frequency of QC prompts is preset to 30 days or 100 tests between QC,
whichever comes first.
Note: The Clarity Platinum analyzer will automatically lock the test strip
function after 30 days or 100 tests have been performed to ensure the
integrity of the urinalysis strips. In order to use the Clarity Platinum analyzer,
CLIA Waived users are required to perform a successful QC test.
Significance of QC tests and Consequences of Not Performing
QC Procedures
Quality control testing is a process of detecting errors within the testing site to
ensure both the reliability and accuracy of test results in order to provide the
best possible patient care. Therefore, to ensure the accuracy and performance
of the Clarity Platinum Test System, quality control procedures are incorporated
into the Clarity Platinum analyzer. The purpose of including quality control
procedures is to evaluate the reliability of the test system by assaying stable
quality control material that resembles patient samples.
Page 16
Clarity Platinum Urine Analyzer
OPERATOR'S MANUAL
button on the main menu screen to access the advanced
button to access
We Support Better Outcomes™

Advertisement

Table of Contents
loading

Table of Contents