Performing A Urinalysis Test; Viewing The Urinalysis Test Result - Clarity Platinum Operator's Manual

Urine analyzer
Table of Contents

Advertisement

Performing a Urinalysis Test

1.
Dip the strip into the urine sample and remove it immediately.
Ensure that all pads on the strip are saturated with the urine sample.
While you are doing this, press the Start button.
2.
After pressing the Start button, the analyzer begins 10 seconds of
self-calibration. During this time, touch the long edge of the strip onto
a paper towel to remove the excess urine, and place the strip on the
strip bed with the test pads facing up. Ensure that the test strip is
inserted fully into the back of the strip bed. The dipping step should
not exceed 10 seconds.
3.
When calibration is complete, the analyzer retracts the strip bed
with the strip, and it will begin analysis.
4.
A timer on screen will count down the time remaining in the strip
analysis process.
Note: You cannot cancel a test before the analyzer finishes the test.
Caution:
DO NOT push or pull the strip bed or move the analyzer
during the test. It might cause the analyzer to fail calibration.

Viewing the Urinalysis Test Result

1.
After the countdown ends, the analyzer displays the test result on
the screen.
2.
Press the Results button to access the stored results, or press the
Next
Test button to perform another test.
Page 10
Clarity Platinum Urine Analyzer
OPERATOR'S MANUAL
We Support Better Outcomes™

Advertisement

Table of Contents
loading

Table of Contents