Operations - Clarity Platinum Operator's Manual

Urine analyzer
Table of Contents

Advertisement

2. Operations

By pressing the
Test Strip
a urinalysis test.
Note: The Barcode Reader feature has been enabled (default configuration)
and you will be prompted to scan or enter the barcode on the test strip bottle
prior to reaching the Test Preparation screen.
Entering Strip Barcode Information
To enter the Strip Barcode information, perform the following steps:
1.
After pressing the
the Clarity Platinum barcode scanner to enter the barcode from the
CLA-10P bottle.
2. After entering the barcode, press Enter
Note: Test known positive and negative specimens/controls at each of the
following events in accordance with local, state, and/or federal regulations or
accreditation requirements. For CLIA Waived settings, Clarity Urinalysis Quality
Controls must be tested under the following conditions:
When a new canister of strips is opened OR
Test results appear to be inaccurate OR
A new operator uses the analyzer OR
Each new day of testing OR
After performing maintenance or service on the analyzer.
(See Section 4.2: Advanced Settings, Quality Control)
Page 8
Clarity Platinum Urine Analyzer
OPERATOR'S MANUAL
button on the main menu screen, you can perform
Test
Strip button, use the keypad on the screen or
to proceed to the next step.