Clarity Platinum Urine Analyzer
OPERATOR'S MANUAL
12.
After pressing the Start
of self-calibration. During this time, touch the long edge of the strip to
a paper towel to remove the excess quality control sample and place
the strip on the strip bed with the test pads facing up. Ensure that the
test strip is fully inserted into the back of the strip bed.
13. When calibration is complete, the analyzer retracts the strip bed
and starts analyzing the strip.
14. A timer on the screen counts down the time remaining in the strip
analysis process.
15. Once the test is completed, the screen will display and print the
QC test result for Control Level II. The QC Pass/Fail message will be
displayed with the test result on the display screen.
16. Press to return to the QC main menu.
Note: Water should not be used as a QC Level I low/negative control solution.
Note: Once the QC test is successfully performed, the test system will display
the results on the screen. Passing the QC test enables the user to conduct
patient sample testing. Failing the QC test will lock the system, and the user
cannot perform any patient testing until the Quality Control test is performed
successfully. If the problem persists, stop the test and call technical support
at (877) 485-7877, Monday-Friday, 8am-6pm EST, or email us at
techsupport@claritydiagnostics.com.
Note: In CLIA Waived settings, QC Prompts are enabled and required. The
analyzer has to pass Quality Control testing before being able to resume
patient urinalysis testing.