Table 7: Electromagnetic Immunity - Nellcor OxiMax NPB-40 Service Manual

Handheld pulse oximeter
Hide thumbs Also See for OxiMax NPB-40:
Table of Contents

Advertisement

Specifications
72

Table 7: Electromagnetic Immunity

The NPB-40 is intended for use in the electromagnetic environment
specified below. The customer or user of the NPB-40 should assure that
it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic
+ 6 kV contact
Discharge (ESD)
+ 8 kV air
IEC 61000-4-2
Electrical fast
+ 2 kV power
transient/burst
supply lines
IEC 61000-4-4
+ 1 kV for
input/output
lines
Power
3 A/m
Frequency
(50/60 Hz)
magnetic field
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
Recommended Separation Distance
3.5
------ -
d
P
=
V 1
Compliance
Electromagnetic
Level
Environment
Guidance
Complies
Floors should be
wood, concrete, or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Complies
Main power should
be that of a typical
commercial or
hospital
environment.
Complies
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
3 Vrms
Portable and mobile
RF communications
equipment should
be used no closer to
any part of the
3 Vrms
NPB-40, including
the cables, than the
recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).