Service Manual for Sunflower Mini-Shaker 3D Periodical maintenance 3D does not require periodical technical maintenance, except cleaning. Cleaning. Standard ethanol (75%) or any other cleaning liquids can be used for cleaning and disinfecting of the device, if it does not contain any strong organic solvents (such as acetone, toluene or benzene).
Page 2
Spare part replacement Disassembling for: Round plate with bearing, Rubber collar 1. Disconnect the unit from power supply. 2. Take off the platform. 3. Unscrew two fixing screws (fig.1/1) on the base platform and remove the base platform. 1. To replace the Round plate with bearing: –...
Page 3
5. To replace the Potentiometer: – unscrew the three screws (fig.5/1) and turn upside down the steel plate; – disconnect the Potentiometer terminal (fig.6/8) from the driver board; – pull off potentiometer knob (fig.5/2) – unscrew the screw (fig.6/3) and release Potentiometer base plate;...
Page 4
3D Service Manual - V.4AW Version 1 - October 20 1 Page 4...