EU Declaration of conformity .................. 15 About this edition of user instructions The current edition of the user instructions applies to the following models and ver- sions of laboratory centrifuges: LMC-3000 ................versions V.6AD, V.6AE LMC-4200R ..................version V.4AD...
Safety precautions The following symbols mean: Caution! Make sure you have fully read and understood the present Manual before using the equipment. Please pay special attention to sections marked by this symbol. GENERAL SAFETY The protection provided can be ineffective if the operation of the appliance does not comply with the manufacturer's requirements.
Page 4
DURING OPERATION Do not centrifuge flammable or chemically active substances. If such liquids are spilled on the rotor or rotor chamber, the centrifuge must be cleaned with a moist cloth and a mild soap solution. Do not use rotors with visible signs of corrosion, wear or mechanical damage. ...
After non-stop centrifugation for 1.5 hours, temperature in the working chamber of the LMC-3000 model can rise, but not by more than 15°C higher than ambient. In case of centrifugation of thermally sensitive samples, we recommend using LMC-4200R model, la- boratory centrifuge with cooling function.
The warranty does not cover in-transit damage. Warranty covers only the units transported in the original package. Caution! Due to the high weight of the LMC-4200R unit, its unpacking and in- stalling must be carried out by two persons. 4.2.
Page 7
According to EN 61010-2-20, people and hazardous materials must not be within a 300 mm area around the device during the centrifuge operation. (LMC-4200R) Be sure to cork the drain hose (fig. 2/1) on the left side of the unit with a stopper to prevent the rising of chamber temperature.
Check the power cable for any signs of damage. Connect the power cable to a properly grounded mains socket. Set the power switch on the side to position I (on). Caution! In model LMC-4200R, cooling system requires some time for start- up. After powering up, wait for about 4 minutes before centrifugation. 5.2.
Page 9
Values convert after changing units. RCF depends on selected rotor or adapters, as shown in Table 1. 5.4.5. Chamber temperature (fig. 5/7, only for LMC-4200R). Select the chamber tempera- ture, step 1 °C. Cooling starts independently from centrifugation.
Page 10
5.12.3. LMC-4200R. Insert the emergency opening tool into the slot on the right side of the unit, above the power switch (fig. 3/1). Push until a click is heard and open the un-...
80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C. Biosan is committed to a continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice.
V.4AD 230 V, 50 Hz BS-010212-AAA 7.2. To inquire about or order the optional accessories or the replacement parts, contact Biosan or your local Biosan representative. 7.2.1. Optional accessories. Rotors and rotor holder. Maximum speed / RCF Catalogue num- Model...
For the ease of cleaning, chamber is equipped with a drain hole with tube (fig. 2/1). 8.3.4. (For LMC-4200R) Clean the condenser of the unit at least once per year. Disconnect from electric circuit. Unscrew 4 screws and remove the grate on the rear panel of the unit, then vacuum the condenser.
Extended warranty. For LMC-4200R, the Premium class model, one year of extended warranty is availa- ble free of charge after registration, during 6 months from the date of sale. Online registration form can be found in section Warranty registration on our website at the link below.
10. EU Declaration of conformity Unit type Mini-centrifuge, laboratory centrifuges Models Microspin 12, LMC-3000, LMC-4200R Serial number 14 digits styled XXXXXXYYMMZZZZ, where XXXXXX is model code, YY and MM – year and month of production, ZZZZ – unit number. Manufacturer SIA BIOSAN Latvia, LV-1067, Riga, Ratsupites str.
Page 16
Biosan SIA Ratsupites 7, build.2, Riga, LV-1067, Latvia Phone: +371 67426137 Fax: +371 67428101 http://www.biosan.lv Edition 4.-6.02 – December of 2018...