CLOCK CALIBRATION
Trinity can be calibrated periodically by connecting it to the 10M/MX Atomic Clock. This is
not something that necessarily needs to be done very often, but if desired, here is how to
do it:
1. Connect 10M to Trinity as normal.
2. Power on both devices and wait for 10M/MX to reach operating temperature.
3. Set the Trinity SOURCE knob (3) to OVEN, and be sure that the ATOMIC LED (14) is lit.
4. Wait one minute to let the units stabilize. If the middle display line is stable at +/0
0.5ppm or greater, you should proceed with the calibration. If not, wait until the display
stabilizes before deciding whether to proceed with the calibration or not.
5. Press the Display button (15) and then the Varispeed button (10). Trinity is now in
Calibration mode.
6. To proceed with calibration, press the SAMPLE RATE knob (4). Trinity is now calibrated
to an accuracy greater than +/- 0.001ppm.
7. Exit Calibration Mode by pressing the INFO button (15).