  • Page 1 Care instructions KaVo uniQa...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    1 User instructions ....................... 2 Reprocessing steps in accordance with ISO 17664 ............2.1 Notes ........................... 7 2.2 General instructions for the reprocessing of KaVo treatment centres ........8 2.2.1 Overview of reprocessing .................. 10 2.3 Reprocessing of the surfaces of treatment centre / upholstery / foot control ......11 2.4 Preparation of the dentist element...................
  • Page 4 Table of contents 2.11 Reprocessing of the amalgam separator................45 2.12 Reprocessing KaVo Screen One / KaVo Screen HD ............46 2.13 Reprocessing of the Service table 1568 and tray holders............ 46 2.14 Reprocessing of the water bottle ..................47 3 Replacing and checking cleaning and germ reduction agents ..........
  • Page 5: User Instructions

    Read these instructions prior to first startup to avoid misuse and prevent damage. Comply with the Instructions for Use of the treatment centre. Please note also the corresponding hygiene video on the KaVo website. Please use the QR code or the following link: https://www.kavo.com/hygiene-easy-way...
  • Page 6 Care instructions KaVo uniQa 1 User instructions Please refer to the serial number of the product in all inquiries! For further in- formation, please visit: www.kavo.com Target group The instructions for use are intended for medical professionals, in particular dentists and dental practice personnel.
  • Page 7: Reprocessing Steps In Accordance With Iso 17664

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.1 Notes 2 Reprocessing steps in accordance with ISO 17664 2.1 Notes The reprocessing consists of the following basic steps: ▪ Cleaning and disinfection ( manually or automatically) ▪...
  • Page 8: General Instructions For The Reprocessing Of Kavo Treatment Centres

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treat- ment centres NOTE The reprocessing methods for the instruments and motors are evident from the separate instructions for use enclosed with the instrument and motor packages.
  • Page 9 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treat- ment centres Manual disinfection 4 Perform wipe disinfection only, the use of spray disin- fection and disinfection baths is not feasible.
  • Page 10: Overview Of Reprocessing

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treat- ment centres ▪ Autoclave with gravity procedure: at least 10 minutes at 134 °C -1 °C/ +4 °C (273 °F -1.6 °F/ +7.4 °F) 4 Use according to the manufacturer's Instructions for Use.
  • Page 11: Reprocessing Of The Surfaces Of Treatment Centre / Upholstery / Foot Control

    Instructions for use of the treatment centre Locking the touch screen 4 Tap the "Lock screen" button on the dentist element or the KaVo Logo on the assistant element to lock the screen for wiping with a disinfectant. ð The following display is shown on the dentist element:...
  • Page 12 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.3 Reprocessing of the surfaces of treatment centre / uphol- stery / foot control CAUTION Product damage due to improper disinfection. Malfunctions. 4 Use the disinfectant according to the specifications of the manufacturer.
  • Page 13: Preparation Of The Dentist Element

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.4 Preparation of the dentist element 2.4 Preparation of the dentist element 2.4.1 Reprocessing of the handle and holder pad / non- slip mat Cleaning and disinfection 4 Press the push-button and pull off the handle (S-table and T-table only).
  • Page 14: Reprocessing Of The Swing Arm S Table

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.4 Preparation of the dentist element 4 Also refer to: Chapter on Automatic cleaning programmes 4 Pull the latches outward evenly and carefully remove the handpiece holder.
  • Page 15: Reprocessing Of The Instrument Hoses

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.4 Preparation of the dentist element 4 Pull-off the swing arm in an upward direction. 4 Clean the swing arm and roller manually. 4 Subject the swing arm S table to wipe disinfection.
  • Page 16: Reprocessing Of Cameras

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.5 Reprocessing of the assembly kit for physiological saline solution and surgical motor 4 If necessary, replace the turbine return air filter (Mat. no. 0.200.3098) and clean the filter holder.
  • Page 17: Automated Reprocessing

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.5 Reprocessing of the assembly kit for physiological saline solution and surgical motor Manual drying NOTE Coolant hose/NaCl hose with accessories is intended for single use only and is not to be disinfected and sterilised.
  • Page 18: Sterilisation By Moist Heat In Accordance With Iso 17665-1 In A Steam Steriliser

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.6 Reprocessing of the assistant element 4 Seal the handpiece holder and motor cable in packaging for sterile items. 2.5.4 Sterilisation by moist heat in accordance with ISO...
  • Page 19: Reprocessing Of The Suction Hoses

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.6 Reprocessing of the assistant element 4 Also refer to: Chapter on Automatic cleaning programmes 4 Pull the latches evenly outwards and carefully remove the handpiece holder.
  • Page 20: Reprocessing Of The Sieve Inserts And Suction Hoses

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional handpiece Cleaning and disinfection of the suction hose connector 4 The manual external cleaning and disinfection of the connector involves manual cleaning of the exterior and wipe disinfection.
  • Page 21: Preparations At The Site Of Use

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional handpiece 2.7.1 Preparations at the site of use WARNING Hazard from contaminated products. Contaminated products are associated with an infection hazard.
  • Page 22: Manual Reprocessing

    Validated manual internal cleaning (removal of residual protein) can be done with KaVo CLEANspray. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the cleaning adapter manual 3F/MF syringe cannula (Mat. no. 3.005.7780). 4 Hold the can vertically.
  • Page 23 KaVo with respect to the compatibility of materials (e.g. WL-cid / made by ALPRO). 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the cleaning adapter manual 3F/MF syringe cannula (Mat. no. 3.005.7780).
  • Page 24: Automated Reprocessing

    2 Reprocessing steps in accordance with ISO 17664 | 2.7 Reprocessing the three-function and multifunctional handpiece 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the cleaning adapter manual 3F/MF syringe cannula (Mat. no. 3.005.7780).
  • Page 25: Care Products And Systems - Servicing

    NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 4 Remove any residual liquids through subsequent sterilisation.
  • Page 26: Reprocessing Of The Patient Element

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.8 Reprocessing of the patient element 2.8 Reprocessing of the patient element 2.8.1 Reprocessing of the tumbler 4 Reprocess as specified by the manufacturer. 2.8.2 Reprocessing of the tumbler holder and tumbler...
  • Page 27: Reprocessing Of The Spittoon Bowl

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.8 Reprocessing of the patient element 2.8.3 Reprocessing of the spittoon bowl CAUTION Water in the device. Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl flush"...
  • Page 28: Automatic Cleaning Programmes

    During routine daily operation of the treatment centre, the continuous germ re- duction function secures continuous germ reduction in the water-conducting systems due to the automatic dosing of the germ reduction agent, KaVo OXY- GENAL 6.* *In combination with DVGW water block only 2.9.1 Preparation of the cleaning programmes...
  • Page 29 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes 4 Remove the cover of the hygiene centre and put it down. 4 Tap the button of the desired cleaning programme. ð The individual preparation steps are shown on the touch screen and can be run in sequence.
  • Page 30 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes 4 Grasp the three-function or multifunctional handpiece on the adapter between hose and handpiece and take off the cannula ② and the gripping sleeve ①.
  • Page 31 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes NOTE To improve the hygiene and lifespan, it is recommended to place the suction hoses over the arm system of the assistant element during the cleaning pro- cess.
  • Page 32: Cleaning Programme, "After Treatment

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes S-Table 4 Lock the swing arm into the front catching position to ensure that the hand- piece is removed. 4 In the case of swing arms without a catching function, insert the swing arm lock (Mat.
  • Page 33: Cleaning Programmes "Morning" And "Evening

    20 seconds each. In addition, the suction system and the spittoon bowl are cleaned and disinfected with DEKA- SEPTOL Gel. KaVo recommends running the "Evening" cleaning programme after surgical interventions. Tipp: Place the tumbler on the tumbler holder first to prevent the water from splashing.
  • Page 34: Setting The Rinsing Time

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes 4 After completion of the "Evening" cleaning program, return the instrument and suction hoses to their holders. 4 After completion of the "Evening" cleaning programme, do not place hoses in holders and switch off the treatment unit.
  • Page 35: Weekly" Cleaning Programme

    The "Weekly" cleaning programme prevents the formation of microorganisms during extended downtimes. During the "Weekly" cleaning programme, the af- ferent water ducts are automatically filled with a higher concentration of KaVo OXYGENAL 6. The "Weekly" cleaning programme takes approx. 45 minutes to complete.
  • Page 36 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes NOTE If a malfunction occurs in the amalgamate separator during "Weekly" clean- ing, the workflow is discontinued. Run the "Weekly" cleaning programme: ▪ At initial start-up ▪...
  • Page 37 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes 4 Wait for the cleaning programme to be completed. 4 Or switch the treatment centre off. ð A progress bar indicates the cleaning steps.
  • Page 38: Discontinuing A Cleaning Programme

    ð The unit is now ready for use. Additional intensive germ reduction during the down- time KaVo recommends to run an additional "Weekly" cleaning, e.g. during semester breaks at universities. This flushes the components and replaces the germ re- duction liquid (after 4 weeks).
  • Page 39: Use Of The Cleaning Programmes With The Water Bottle

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes NOTE In the "Weekly" cleaning programme, a higher concentration of OXYGENAL 6 is present in the treatment centre. If this programme gets discontinued, the water-conducting systems are first rinsed with the lower concentration before the treatment centre can be reused.
  • Page 40 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes Preparation of water bottle with standard OXYGENAL 6 concentration Confirm the completion of the preparation on the dentist element. 4 Press the "Yes" button to confirm that the water bottle containing the re- quired concentration was attached.
  • Page 41 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes Cleaning programme, "Weekly" The "Weekly" cleaning programme runs an intensive germ reduction at a higher OXYGENAL 6 concentration. The following amount of OXYGENAL 6 is required:...
  • Page 42 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes 4 Subsequently, the "Continue" button is shown. Press the "Continue" button to start the cleaning programme. 4 After completion of the exposure phase, remove the remaining water from the water bottle.
  • Page 43 Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes Preparation of water bottle with standard OXYGENAL 6 concentration Confirm the completion of the preparation on the dentist element. 4 Press the "Yes" button to confirm that the water bottle containing the re- quired concentration was attached.
  • Page 44: Reprocessing Of The Hygiene Centre

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.10 Reprocessing of the hygiene centre ð The cleaning programme proceeds automatically. 4 After completion of the "Weekly" cleaning programme, place hoses in hold- ers. 2.10 Reprocessing of the hygiene centre CAUTION Damage caused by inappropriate cleaning agents.
  • Page 45: Reprocessing Of The Amalgam Separator

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.11 Reprocessing of the amalgam separator 4 Check the drain. 4 Clean the handpiece holder of the hygiene centre insert under running water and insert it. 4 The handpiece holder of the hygiene centre insert is thermodisinfectable.
  • Page 46: Reprocessing Kavo Screen One / Kavo Screen Hd

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.12 Reprocessing KaVo Screen One / KaVo Screen HD 4 Remove the cover. 4 Replace the amalgam container ① according to the instructions for use of the amalgam separator.
  • Page 47: Reprocessing Of The Water Bottle

    Care instructions KaVo uniQa 2 Reprocessing steps in accordance with ISO 17664 | 2.14 Reprocessing of the water bottle 4 Clean the surfaces with a soft cloth and water, and wipe-disinfect them with an approved surface disinfectant. Norm tray and non-slip mat can be sterilised.
  • Page 48 The water in the water bottle needs to be changed at least once daily. In or- der to ensure the efficiency of the measures, KaVo urgently recommends re- processing the water with KaVo OXYGENAL 6. Also refer to: Care instructions of the treatment centre.
  • Page 49: Replacing And Checking Cleaning And Germ Reduction Agents

    Care instructions KaVo uniQa 3 Replacing and checking cleaning and germ reduction agents | 3.1 Replacing the DEKASEPTOL Gel bottle 3 Replacing and checking cleaning and germ reduction agents CAUTION Working with cleaning, germ reduction and disinfection agents neces- sitates a certain degree of diligence.
  • Page 50: Replacing The Oxygenal 6 Bottle

    NOTE Before you use the OXYGENAL 6, check the expiry date of the product. The only permissible agent for use in KaVo water germ reduction units for wa- ter is KaVo OXYGENAL 6 (Mat. no. 0.489.3451). If the OXYGENAL 6 needs to be replenished, the display of the dentist element shows a status message: "Oxygenal empty"...
  • Page 51 Care instructions KaVo uniQa 3 Replacing and checking cleaning and germ reduction agents | 3.2 Replacing the OXYGENAL 6 bottle 4 Remove the cover from the unit base. 4 Unscrew the lid of the new OXYGENAL 6 bottle. 4 Swing out the empty OXYGENAL 6 bottle from the unit base, and pull it down off the adapter.
  • Page 52: Replenishing Oxygenal 6 When Using The Water Bottle

    4 Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing device. 4 Guide the tube through the orifice of the coarse sieve of the KaVo OXY- GENAL 6 bottle. 4 Screw the KaVo OXYGENAL 6 dosing device tightly to the KaVo OXYGENAL 6 bottle.
  • Page 53: Checking The Oxygenal 6 Concentration

    Care instructions KaVo uniQa 3 Replacing and checking cleaning and germ reduction agents | 3.4 Checking the OXYGENAL 6 concentration 4 Then fill-up the water bottle with tap water. NOTE Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended downtimes after condensa- tion of the water.
  • Page 54 Care instructions KaVo uniQa 3 Replacing and checking cleaning and germ reduction agents | 3.4 Checking the OXYGENAL 6 concentration 4 Test the standard concentration 0.02% H (200ppm): Fill a measuring cup (40 ml) with 39 ml water and 1 ml treated water.
  • Page 55 Care instructions KaVo uniQa 3 Replacing and checking cleaning and germ reduction agents | 3.4 Checking the OXYGENAL 6 concentration Other concentrations can be tested using the following table: Concentration Measuring cup Ratio water : Value shown on treated water test strip 0.02 % (200 ppm)