Kane 79 Manual

Kane 79 Manual

Carbon monoxide monitor and personal co alarm

Advertisement

Quick Links

KANE79
Carbon Monoxide Monitor
and Personal CO Alarm
KANE
LINK
Stock No: MAN00293 Rev: 1.01122
NOVEMBER 2022
©
Kane International Ltd

Advertisement

Table of Contents
loading

Summary of Contents for Kane 79

  • Page 1 KANE79 Carbon Monoxide Monitor and Personal CO Alarm KANE LINK Stock No: MAN00293 Rev: 1.01122 NOVEMBER 2022 © Kane International Ltd...
  • Page 2: Table Of Contents

    VISUAL INDICATION AUDIBLE & VIBRATING ALARMS BATTERY STATUS REPLACING BATTERY REPLACING BATTERY KANE LINK TROUBLESHOOTING SPECIFICATIONS CLEANING COLD WEATHER PRECAUTIONS WHERE TO SEND YOUR ANALYSER EU DECLARATON OF CONFORMITY KANE CARE 18-21 KANE & UKAS RECYLING THANK YOU 2 KANE79 MANUAL...
  • Page 3: Kane79 Overview & Safety Warning

    Data hold • Battery Indication • KANE LINK Wirelessly connect up to 4 KANE79 CO monitors to your KANE LINK • analyser When powered on, they replace or add measurements to your KANE • LINK analyser See page 12 to add, manage or delete your KANE LINK device •...
  • Page 4 SAFETY WARNING! CO is life threatening even at low concentrations. Ensure you understand the risks before investigating This monitor measures combustion gases that may be toxic in relatively low concentrations. This monitor must only be used by trained and competent persons after due consideration of all the potential hazards.
  • Page 5: Kane79 Features And Keypad

    KANE79 FEATURES AND KEYPAD Status LEDs Sensor cover Battery Large life display Max reading Data hold button button Power On/Off button Belt clip and serial number KANE79 MANUAL 5...
  • Page 6: Operating Instructions

    OPERATING INSTRUCTIONS Every button press makes an audile beep WARNING! Always power on in outdoor fresh air away from boiler flues, vehicle exhausts, bbq, smokers and any other sources of CO. 1. Power on by pressing On/Off button 2. Zero countdown starts at 30 3.
  • Page 7 6. Displaying Maximum CO measured Press MAX to display highest measurement while monitor is ON Press again to return to current value Maximum value is reset when monitor powered off & on 7. Freeze or hold display Press to freeze or hold display Press again to return display to current value Hold value is reset when monitor powered off &...
  • Page 8: Measuring Co

    MEASURING CO Note: Ensure monitor sensor cover is free from obstruction. KANE79 is an ambient air monitor This means it will check for spillages around fires or cookers but does not replace tests requiring specific appliance sampling probes specified, for example, in BS7967.
  • Page 9: Visual Indication

    VISUAL INDICATION These values are based on known environmental and health standards: Concentration Inhalation Times and Toxic Symptoms of CO in air 10ppm Maximum allowable concentration for an 8 hour period without any known health risks according to the UK Department of Health and British Standards BS7967 30ppm Maximum allowable concentration for competent persons to...
  • Page 10: Audible & Vibrating Alarms

    AUDIBLE & VIBRATING ALARM Your KANE79 has 3 alarm levels: CO Levels Indication 2-9ppm Green LED flash once every 3 seconds 10-19ppm Red LED flash once every 2 seconds > 20ppm Red LED flash once per second, sound and vibration active BATTERY STATUS Full battery Low battery...
  • Page 11: Replacing Battery

    REPLACING BATTERY KANE79 battery compartment is located underneath the belt clip - Belt clip is spring loaded to prevent rotation in normal use. Belt clip in normal use 1. Lift belt clip at pivot end against the spring by a few millimeters then rotate 90 degree to clear battery compartment cover.
  • Page 12: Kane Link

    Serial number must be 10 digits long - If shorter, enter 0’s to make up to 10 For example - enter serial number J12345678 as 0012345678 Other KANE LINK devices can be paired - Contact KANE for more details 12 KANE79 MANUAL...
  • Page 13: Troubleshooting

    TROUBLESHOOTING If display fails to zero within minutes the sensor has reached end of life or was exposed to excessive levels of CO - Contact KANE if this continues. Always check for correct operation before each use. To test your KANE79, use a known source of CO taking care to act in a safe manner and not expose sensor to extreme temperatures or CO concentrations - Contact KANE if in doubt.
  • Page 14: Specifications

    SPECIFICATIONS Range: 0-999ppm Resolution: 1ppm Accuracy: ±3ppm 0 to 100ppm¹ ±5% reading > 101ppm Audible Alarm: See page 11 Visual Indication: LCD and LED Battery: 9V alkaline PP3 Battery Life: Up to 250 hours² Duracell Procell Ambient Temperature: 0-40ºC (32-104ºF) Warranty: 1 Year ¹...
  • Page 15: Cleaning

    CLEANING Clean this monitor with a lint free damp cloth and a small quantity of non-abrasive detergent. Avoid moisture entering the sensor cover and after cleaning, leave in a warm dry place to evaporate any surface dampness. Do not use solvent cleaner - It may damage the case, display & sensor. KANE79 MANUAL 15...
  • Page 16: Cold Weather Precautions

    If your monitor is affected by condensation or water ingress, leave in a warm place for a few hours. If you still experience problems please contact KANE Customer Service. WHERE TO SEND YOUR ANALYSER - see section Northern Customer Service Southern &...
  • Page 17: Eu Declaraton Of Conformity

    This declaration of conformity is issued under the sole responsibility of the manufacturer:- Kane International Ltd. Kane House, 11 Bessemer Road, Welwyn Garden City, Hertfordshire, AL10 1GF, UK. Tel: + 1707 375550 Web: www.kane.co.uk The KANE79 is in conformity with the relevant Union harmonization legislation below:...
  • Page 18: Kane Care

    Welcome to KANE CARE KANE CARE is our award winning promise to never let you down or your money back Analysers & CO monitors must be serviced and recertified every year - In some markets this is a legal requirement KANE CARE is all inclusive and applies to any KANE analyser or monitor booked online via www.kane.co.uk &...
  • Page 19 +44 1707 375550 Email: sales@kane.co.uk Trustpilot KANE and the Kane logo (registered in the UK and elsewhere) are trademarks of Kane Group Limited, used under licence All other marks are trademarks of their respective holders. © 2022 Kane International Limited...
  • Page 20 Select FREEPOST for tracked carriage - UK & Ireland mainland only Your analyser will be despatched the same day we receive it… OR YOUR MONEY BACK* *Excludes KANE-EGA, AUTO600, ‘9 series’ analyser & UKAS certificates - Ireland 2 day turnaround 20 KANE79 MANUAL...
  • Page 21 International All Inclusive Price KANE CARE Service & Recertification email: sales@kane.co.uk with analyser name & serial number Pay our all inclusive KANE CARE quote Send your analyser & accessories We will return free of charge* after KANE CARE Service & Recertification with 1 year warranty if under 10 years old *Excludes local tax &...
  • Page 22: Kane & Ukas

    UKAS is the only recognized UK authority assessing laboratory standards & capability KANE is the only Flue Gas Analyser manufacturer accredited by UKAS to ISO 17025:2017 for gas certification UKAS calibration certificates from KANE are legal documents you can submit in evidence to a UK court - they prove your analyser...
  • Page 23: Recyling

    RECYCLING THIS PRODUCT CONFORMS TO PLEASE RECYCLE - CONTACT US PACKAGING MADE IN THE UK KANE79 MANUAL 23...
  • Page 24: Thank You

    Thank you for buying this product. Before use, please register on our website www.kane.co.uk Scan QR code to directly register your product online. Kane International Ltd Kane House, 11 Bessemer Road Welwyn Garden City Hertfordshire AL7 1GF, UK email: sales@kane.co.uk...

Table of Contents