3.
As you are powering up the autoclave for the first time, the autoclave will be in a low water
condition, the low water indicator should now be lit and the error code F004 will show in the
temperature display. You may ignore this at this stage as opening the autoclave and filling it
with water will reset this error.
4.
Press the 'door' button on the control panel there will be
a bleep and the message 'Hold' will be displayed in the
temperature display. Wait for a short time until the
temperature display returns to normal, there is another
bleep and the door indicator illuminates. The door button
can now be pressed again to release the lock.
Open the autoclave door as described below:
5.
a) Lift the locking handle up
b) The handle will now spring out into its unlocking
position
c) Lift the handle fully upwards to unlock the door. The
handle is now in its safety lock position, allowing any
residue of pressure inside the autoclave to escape
harmlessly.
d) Move the handle slightly down to release it from the
safety position
e) Push the handle in as far as it will go
f) Move the handle fully upwards to its parked position
g) With the lid unlocked, carefully pull the door open
Door opening procedure
f) Door
unlocked
d) Safety Lock
Position
Safety Lock
Release
Position –
WAIT!
e) Push in
c) Unlock
b) Spring Release
a)
Door Locked
Position