Appendix B - Steam Table - Priorclave Q63 Operation & Maintenance Manual

Autoclaves
Table of Contents

Advertisement

Priorclave Manual 27125322-EN Issue 2

Appendix B - Steam Table

Autoclave
Correct Correlation between Temperature and Pressure shows correct operation of the
autoclave and that air purging is satisfactory.
This table is accurate at sea level and at moderate altitudes will be sufficiently accurate for its
intended purpose. However, at higher altitudes the pressures indicated will be slightly higher
than those shown above.
Temperature
o
o
(
C)
(
F)
100
212
105
221
110
230
115
239
120
248
121
250
122
252
124
255
126
259
128
262
130
266
132
270
134
273
136
277
138
281
140
284
06/06/2012
Pressure (BarG)
0.00
0.20
0.43
0.69
0.99
1.06
1.13
1.25
1.35
1.55
1.70
1.86
2.04
2.21
2.40
2.60