Changing Date & Time - Priorclave Q63 Operation & Maintenance Manual

Autoclaves
Table of Contents

Advertisement

Priorclave Manual 27125322-EN Issue 2
Changing Date & Time
A number of additional control system settings can be accessed via a "Hidden Menu".
To access these settings turn & hold the thermal lock key in the override position. Press the
time up or down keys. Release the thermal lock key. 1 is displayed on the temperature
display, by default 0 will displayed on the time display. The temperature display now shows
the number of a list of operating parameters, the value for the parameter is shown in the time
display. Scroll through the list of available parameters using the temperature up/down keys.
After no keys are pressed for eight seconds the display returns to normal.
The function of these settings is as follows:
Temp.
Time
Display
Display
The autoclave can be set for the cycle to start after a pre-programmed delay, for
example to allow a media preparation cycle to complete shortly prior to the start
of the working day. Setting the value of parameter 1 to 1 in the time display
switches delayed start on.
1
2
3
*
4
5
6
7
8
9
10
#
11
+
The time is set in real time, therefore the clock has to be correctly set for this to work
properly.
After one delayed start operation, delayed start automatically switches off, and the
autoclave returns to normal operation.
#
Models fitted with optional Cycle Repeat Facility only
* Models fitted with 5 or 10 Program Memory
Setting marked * are program number related, and therefore should you wish to use
different values for these in different programs this can be done by changing the value
when the correct program is selected.
Function
0-24
Delayed Start Time Hour
0-59
Delayed Start time Minute
0/1
Start Delay Select On/Off
0-999 Print Interval
Year Setting
Month Setting
Date Setting
Hour Setting
Minute Setting
Second Setting
Scroll back up to parameter 1 to confirm the new or current time
settings.
1-999 Cycle Repeats
06/06/2012
Action
+
Enter required Start time hour (24 hour
clock)
+
Enter required Start time minute
+
1= ON
0= OFF
Enter time (minutes)between printing
during process time (0= printer disabled)
Enter Year
Enter Month
Enter Day of month
Enter Hour
(24 Hr Clock)
Enter Minute
Enter Second.
Enter Number of Cycles Required
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ps/q63/eh320

Table of Contents