a) Move the locking handle down to
release it from its parked position
b) The locking handle will now spring out
c) Move the locking handle down to lock
the autoclave door
d) Push the handle in fully against the
spring
e) Move the handle fully down into its
'park' position
11. Wait a few seconds for the 'start' indicator
to illuminate, and press the 'start' button to
begin the cycle.
During the cycle, check that there are no
12.
problems during heat-up and process.
Following successful completion of the
commissioning cycle your Priorclave is ready
to process its first working load.
Please refer to the Operation section later in this manual before running your first working
load as this gives further details on operation of the autoclave and on the control options
which may be fitted
If you experience any problems during this procedure please contact Priorclave service or your
local agent.

Full Commissioning and Performance Qualification

If you are having the unit commissioned by a Priorclave technician this will be a simple matter
of checking for correct installation, checking that all functions are operating correctly, and
familiarising you with the autoclave. There are, however some benefits that can be gained
from having your Priorclave commissioned to suit your particular loads and requirements.
Some examples of settings that can be optimised during commissioning, and the advantages
these can provide are listed below.
Establishing optimum freesteam temperatures for effective air displacement.
-
Establishing optimum freesteam time for effective air displacement, whilst eliminating
-
unnecessary time and energy consumption.
Establishing optimum process time and temperature to ensure complete sterilisation,
-
whilst maintaining minimum cycle time and energy consumption.
Setting thermal lock release temperature to suit your particular load, to eliminate
-
unnecessary cooling time whilst ensuring safety.
If you feel that any or all of the above would be of use to you then please contact Priorclave
Service.
06/06/2012
Door closing procedure