Cycle Abort And Thermal Lock Override - Priorclave Q63 Operation & Maintenance Manual

Autoclaves
Table of Contents

Advertisement

Priorclave Manual 27125322-EN Issue 2
Move the locking handle down to
a)
release it from its parked position
The locking handle will now spring
b)
out
Move the locking handle down to
c)
lock the autoclave door
Push the handle in fully against the
d)
spring
Move the handle fully down into its
e)
'park' position
Wait a few seconds for the 'Start'
12.
indicator to illuminate and press the
'Start' button to begin the cycle.

Cycle Abort and Thermal Lock Override

Aborting a cycle
To abort the cycle at any stage press the 'Start' Button
Thermal Lock Override
First abort the cycle as above.
After checking that there is no pressure within the autoclave turn the thermal lock key to the
right hold it there.
Press the 'Door' button once, keeping the thermal lock key held over.
Wait during the 'Hold' display until the 'Door' indicator illuminates.
Keep the key held and press the 'Door' button once to unlock the door.
The key-switch can now be released and the door opened as above.
If the key is released at any stage the procedure must be repeated to open the door and reset
the display.
06/06/2012
Door closing procedure
a) Door unlocked
Position
b) Spring
Release
d) Push in
e) Door Locked
Position
c) lock
11