Priorclave Manual 27125322-EN Issue 2
F010
Air detector input activated. If fitted the air detector system has detected an over pressure
condition symptomatic of excess air remaining in the load.
F011
Printer Timeout / Malfunction. The control system has nor received confirmation from the
printer within its pre-set timeout.
F012
Door micro-switch fault. If a door micro-switch opens during a cycle this fault code is displayed
F013
Jacket Timeout -If a jacket is fitted it has not reached the required temperature within the Pre-
set time. This would indicate a problem with steam supply or inlet or drain valve operation
F014
Jacket Over temperature - If a jacket is fitted the temperature has exceeded the pre-set alarm
temperature
F015
Jacket under temperature - If a jacket is fitted the temperature has fallen below the pre-set
operating temperature band.
F016
Water Fill Timeout - The upper level water probe level has not been reached within the
allowed time for filling and the filling operation has been stopped. This function prevents
continuous unsupervised operation of the water fill, which could lead to flooding.
F017
Free-Steam - During Pulsed Free-steaming operation the lower of the two set temperatures
has not been achieved. The temperature has not fallen sufficiently following the opening of the
vent valve.
F018
Heater Overheat. If this fault occurs then the most likely cause is a Low Water condition. Check
the water supply is turned on and the condition of the heater before resetting this fault.
LOCK
This warning will light when the thermal lock keyswitch is in the override position.
LOAD + FAULT CODE F001
This warning is activated in the event of the failure of the load sensing thermocouple. The
thermocouple should be replaced as soon as possible. Great care should be taken to ensure
that loads which would ordinarily be autoclaved with load sensed process timing are
adequately sterilised.
CANCELLING FAULT MESSAGES
The fault messages are cancelled by first correcting the source of the original fault, then
turning the setting lock key switch to position 3. If a key-switch is not fitted they are cancelled
by pressing the reset button.
If 2 or more faults occur at the same time, the one with the highest priority is displayed. (F000
is the highest priority and F012 is the lowest.) If a higher priority fault is cleared it will be
replaced by the next active fault, unless this too is cleared by the same action.