INSTALLATION
5 .

CONNECT TUBE

5 . 1 Prepare Reagent

Prepare the reagents and connection kits shown below.
Reagent Part No.
884-0871-1
974-0521-6

5 . 2 Connect CELLPACK

(1)
Prepare the following reagent tube.
Tube Polyurethane 4 mmID × 6 mmOD : 1 pc
NOTE:
(2)
Connect the diluent (CELLPACK) inlet aspiration nipple behind the unit and the nipple
of the cubitainer spout kit No. 1 with the tube polyurethane 4 mmID × 6 mmOD.
(3)
Set the cubitainer spout kit No. 1 to the CELLPACK container.
A -6
Type
PK-30L
CELLPACK
SWH-200A
STROMATOLYSER-WH
• Cut the tube to an appropriate length, and connect it.
C E
L L
P A
C K
Figure A-5-1: Setting the Cubitainer Spout Kit
Sysmex KX-21 Operator's Manual -- Revised October 1998
Part Name
Cubitainer spout kit No. 1 (for 20L)
Float switch No. 23
Cap
Cubitainer Spout Kit No. 1
Connection Kit