Mode of operation
Ingress protection
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables.
Only use this Philips Avent single/double electric breast pump Premium Plus in combination with the
adapter supplied.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Premium Plus is intended for use in the
electromagnetic environment listed below. The user of the device should ensure that it is used in such an
environment to prevent improper operation.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Single electric breast pump
Continuous operation
IP22 (motor unit), IP20 (adapter)
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Immunity test level
±8 kV contact
±15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Double electric breast pump
Non-continuous operation
55 min on, 180 min off at 22 °C (72 °F)
(duty cycle 23 %)
20 min on, 90 min off at 40 °C (104 °F)
(duty cycle 18 %)
Electromagnetic environment –
guidance
The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
Electromagnetic environment -
guidance
Floors can be a source of static electricity. When
the air is very dry (<20 % RH) the user is advised
to prefer to use the device in a room with
concrete or wooden floors (instead of using it in
a room with a synthetic floor).
English
27