7.5. Additional Operations

Viewing the set value for RPM during shaking
Set the unit to RPM mode

Turn the encoder ONE CLICK right (+) or left (-). On the display the RPM light will blink for 2 sec. and will display the set RPM. After

2 sec. the display will show the actual RPM and the light will stop blinking.
Viewing the set value for Time during shaking
Set the unit to Time Mode

Turn the encoder ONE CLICK right (+) or left (-). The Time light will blink for 2 sec. and will display the set Time. After 2 sec. the

display will show the actual Time and the light will stop blinking.
Viewing the set value for Temperature during shaking
Set the unit to Temperature Mode

Turn the encoder ONE CLICK right (+) or left (-). The Temperature light will blink for 2 sec. and will display the set Temperature. After

2 sec. the display will show the actual Temperature and the light will stop blinking.

8.0 Troubleshooting

The Labnet VorTemp™ 1550 shaking incubator features built in self-diagnostic procedures which are constantly checking the
operating parameters and performance, as well as functions that are necessary for safe and reliable operation. An error code is shown
on the LCD display if an error occurs.
Sample of Error display:
E 1.1: Something is wrong with the motor regulation (PWM regulator, pulse generator, motor). The unit will automatically stop.
Call for service.
E 1.2: The motor does not reach the set RPM in 30 sec. The unit will automatically stop. Call for service.
E 1.3: The set RPM oscillates more than 100 RPM in 2 seconds. The unit will automatically stop. Call for service.
E 2.1: The temperature sensor is not working properly with regards to the heater. The unit will automatically stop. Call for service.
E 2.2: The temperature sensor registers a 5°C higher temperature than the set temperature. The unit will automatically stop. Call for
service.
Er 5:
The lid of the shaker is opened for more than 5 seconds. The unit will automatically stop. To restart the shaker, close the lid
and press Start.
When shipping your equipment for service, follow the packing guidelines listed below:
Clean the inside of the chamber and platforms according to GLP (Good Laboratory Practices) standards, especially if you have used

the equipment with hazardous biological or radioactive materials.
A written description of the error should accompany the unit.

Use the original shipping container and packaging materials if possible.

Run
Time [min]
Power
RPM [x10]
Temp [°C]
Encoder
Er5
Start
Stop
Instruction Manual
|