2.1 Hold Register DO

If applicable, in the Value/description field the Factory Default setting (D), Low limit setting (L) and
High limit setting (H) are explained. NaN does mean that there is Not a Number defined.
NOTE
Because no limit checks are implemented, incorrect settings or settings outside the specified range
are accepted as is. Please take care that settings are done correctly and within the specified range.
Item name
Address
1031
Firmware version
2089
Oxygen unit
2091
Oxygen value
2093
Error
2409
Temperature Unit
2411
Temperature Value
All Rights Reserved. Copyright © 2022, Yokogawa Process Analyzers Europe B.V.
MODBUS Communication – Hold Register DO
Data type Access Value / description
ASCII
R
Uint32
R/W
Float
R
R
Uint32
Uint32
R/W
Float
R
Alphanumeric Firmware version code:
16 bytes, 8 addresses
0x 0000 0010
% vol O
2
0x 0000 0020
% air saturation
0x 0000 0040
ppb (µg/L)
0x 0000 0080
ppm (mg/L)
2)
0x 0400 0000
µmol/L
0x 0800 0000
% DO saturation
0x 2000 0000
hPa (mbar)
1) 2)
0x 4000 0000
ppm gas
0x 8000 0000
Torr
Remark:
1)
ppm gas only possible with DO72-T
2)
Setting not possible in
measurement mode "Dry"
(register HR5703)
D: % DO saturation
Calculated oxygen value in oxygen unit
as defined in register HR2089
Remark:
1)
The calculated oxygen value is only
updated if an interval rate is set
(register HR3499)
2)
When value is -999 this means
there is no calculated output
For details see Section 3.1
0x 0000 0002
K
0x 0000 0004
°C
0x 0000 0008
°F
D: °C
Measured temperature value used for
temperature compensation
TI 12J06J01-00EN-P