Advertisement

Address
Item name
5157
Oxygen Calibration Low Point Status
Oxygen Unit for Oxygen Calibration Low
5159
Point
5161
Activate Oxygen Calibration Low Point
5181
Low Calibration Point Timestamp
5189
Oxygen Calibration High Point Status
5193
Activate Oxygen Calibration High Point
5213
High Calibration Point Timestamp
All Rights Reserved. Copyright © 2022, Yokogawa Process Analyzers Europe B.V.
MODBUS Communication – Hold Register DO
Data type Access Value / description
Uint32
R
Uint32
R
Float
R/W
Uint32
R
Uint32
R
Float
R/W
Uint32
R
For details see Section 4
0x 0800 0000
% DO saturation
Remark:
1)
Shows the oxygen unit for the
oxygen calibration low point.
D: % DO saturation
Write any value to trigger a calibration
for the low oxygen calibration point.
If no calibration error regarding the low
point (bits 2 to 7, see Table 4.1)
occurred, the measured phase and
temperature values (or the manually set
temperature value) will be written in the
registers for Cal0 (register HR5521) and
T0 (register HR5523)
The timestamp the last low point
calibration point was performed
For details see Section 4
nd
Write the desired O
-2
value to trigger
2
a calibration for the high oxygen
calibration point. The unit is set in
HR5535.
If no calibration error regarding the high
point (bits 10 to 15, see Table 4.1)
occurred, the measured phase,
pressure, and temperature values (or
the manually set temperature value) will
be written in the registers for Cal-2
nd
(register HR5529), T2
(register
HR5531) and pATM (register HR5533)
The timestamp the last high point
calibration point was performed
TI 12J06J01-00EN-P
14