3. Failures and Warnings DO

Device failures and Measurement warnings are checked and identified by bit content (0 or 1) in
specific Hold Registers (HR2093 and HR2094). The definition of each bit, the root cause and
possible remedy are explained in table 3.1.
Table 3.1: Bit assignment for DO71/DO72 failures/warnings
Bit#
Bit definition
0:
0x00000000
Reference channel overflow
1:
0x00000002
Temperature too high
2:
0x00000004
Reference photo detector overflow A
3:
0x00000008
Signal channel overflow
4:
0x00000010
Reserved
5:
0x00000020
Reserved
6:
0x00000040
No sensor calculation /
Amplitude too low
7:
0x00000080
Pulse counter overflow
8:
0x00000100
Reference amplitude out of range
9:
0x00000200
Signal photo detector overflow
10:
0x00000400
Reference photo detector overflow B
11:
0x00000800
Memory Write error detected
12:
0x00001000
PME timeout
All Rights Reserved. Copyright © 2022, Yokogawa Process Analyzers Europe B.V.
MODBUS Communication – Failures and warnings DO
Root cause & Remedy
Reference amplitude too high.
Decrease reference current.
The temperature is too high for
performing a measurement.
Reference amplitude is too
low/high;
Sensor cap is not attached;
Too much ambient light.
Adjust reference current;
Attach sensor cap;
Reduce ambient light.
Signal amplitude too high.
Decrease signal current.
Signal amplitude is too low;
Sensor cap is not attached.
Decrease signal current;
Attach sensor cap.
Pulse counter reached its
maximum value.
Reference amplitude is too
low/high;
Sensor cap is not attached.
Adjust reference current;
Attach sensor cap.
Signal amplitude is too high;
Sensor cap is not attached;
Too much ambient light.
Adjust signal current;
Attach sensor cap;
Reduce ambient light.
Reference amplitude is too
low/high;
Sensor cap is not attached;
Too much ambient light.
Adjust reference current;
Attach sensor cap;
Reduce ambient light.
Memory errors occurred.
Contact the relevant
YOKOGAWA sales office.
Internal error.
Contact the relevant
YOKOGAWA sales office.