Basic Setup with the device display

5.4.1 Function "measured value" (000)
ENDRESS + HAUSER
+
E
This function displays the current measured value in the selected unit
(see "customer unit" (042) function). The number of digits after decimal point can be selected
in the "no.of decimals" (095) funtion.
5.4.2 Function group "basic setup" (00)
ENDRESS + HAUSER
+
E
5.4.3 Function "tank properties" (002)
ENDRESS + HAUSER
+
E
This function is used to select the tank properties.
Selection:
• standard
• aluminium tank
• plastic tank
• bypass / pipe
• Coax probe
• concrete wall
standard
The "standard" option is recommended for normal containers for rod and rope probes.
aluminium tank
The "aluminium tank" option is designed especially for high aluminium silos that cause an
increased level of noise when empty. This option is only useful for probes longer than (<4 m).
For short probes (<4 m) select the "standard" option!
26
Levelflex M FMP41C 4...20 mA HART
Endress+Hauser