2.0 Safety Information Before using the Labnet Prism R refrigerated microcentrifuge for the first time, please read this entire manual carefully. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Indoor use only Altitude up to 2,000 meters Pollution Degree 2 4.0 Package Contents Labnet Prism R refrigerated microcentrifuge Rotor with lid, 24 x 1.5/2 mL Rotor removal tool Electric fuses (7A and 3.15A) ...
Securing screw Figure 1. Loading the rotor to achieve balance. To replace the rotor, first make sure the motor shaft and rotor mounting hole are clean. Place the rotor on the motor shaft. Reinstall the rotor securing screw on the motor shaft by turning it clockwise. Hold the rotor with one hand and tighten the rotor securing screw, using the rotor wrench.
They can also be disabled if desired by pressing the Time button six times. The audible signal function can be resumed by pressing the Time button six times again. Definition of the audible sounds: 2 beeps: Power on machine 3 beeps: End of a run LABNET PRISM R REFRIGERATED MICROCENTRIFUGE...
8.0 Operation CAUTION: Never attempt to operate the centrifuge with rotors or adapters that show signs of corrosion or mechanical damage. Never centrifuge strongly corrosive materials that may damage the rotors, accessories, or bowl of the unit. 8.1 Attaching Rotor Lid After the rotor has been properly secured and loaded, attach the rotor lid to the rotor.
The current value flashes, and the new values can be input. Once the display switches into the current value mode after completion of the entered values, the new values are activated. The new values are stored and executed by pressing any function button again or by waiting 6 seconds. LABNET PRISM R REFRIGERATED MICROCENTRIFUGE...
Should you have a question about the operation of the Labnet Prism™ R refrigerated microcentrifuge or if service is required, contact Corning Customer Service. Do not send in a unit for service without first calling to obtain a repair authorization number. Should the unit require return for service, it should be properly packed to avoid damage.
The RCF value is the point where the line crosses the center column. The nomogram is based on the formula: RCF = (11/17 x 10 ) RN where: R = Radius in mm from centrifuge spindle to point in tube bottom N = Speed of spindle in RPM LABNET PRISM R REFRIGERATED MICROCENTRIFUGE...
Claims for transit damage should be filed with the transportation carrier. In the event this product fails within the specified period of time because of a defect in material or workmanship, contact Corning Customer Service at: USA/Canada 1.800.492.1110, outside the U.S. +1.978.442.2200, visit www.corning.com/lifesciences, or contact your local support office.
Page 12
Not for use in humans. These products are not intended to mitigate the presence of microorganisms on surfaces or in the environment, where such organisms can be deleterious to humans or the environment. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications.