For the complete Corning Lambda EliteMax Benchtop Pipettor Instruction
Manual (CLS-AN-671DOC), visit www.corning.com/lifesciences.
Instrument Identification
Liquid Head
Touch Screen
Controller
Instrument Deck
Manually
Operated
Tip Ejector
Single-
channel
Head
Unboxing
Tools needed: Phillips and flat head screwdrivers, 2 mm Allen
wrench (provided), hammer, and needle-nose pliers.
Pry open top of box with the flat head screwdriver.
Remove the screws holding down the deck plate with the
Phillips head screwdriver.
Remove the unit from the box, lifting from the bottom of the
deck plate, and place on the benchtop.
Be sure to remove the holding bracket prior to use. The holding
bracket is located next to the yellow warning label and can be
removed using the Allen wrench provided.
EliteMax
Disposable Tips and
Plates (not included)
Multi-
channel
Head
Installation
4
1
2
1. Connect the Touch Screen Controller cable into the rear panel
of the Corning Lambda EliteMax Benchtop Pipettor.
2. Connect the round connector of the universal external power
supply to the rear panel of the instrument. Push in firmly until
you feel it click into position.
3. Connect the AC power cord to the universal external power
supply and then to the AC wall power receptacle. The AC
power receptacle should be able to supply approx. 300W and
MUST be grounded to a true earth ground.
4. Power up the unit by rocking the Reset/Off switch located in
the back of the Corning Lambda EliteMax Benchtop Pipettor
to the Reset position. The instrument should boot up, and the
home screen will appear in a few seconds.

Advertisement

loading
Need help?

Need help?

Do you have a question about the Lambda EliteMax and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for CORNING Lambda EliteMax

  • Page 1 Corning ® Lambda ™ EliteMax Semi-automated Benchtop Pipettor Quick Start Guide For the complete Corning Lambda EliteMax Benchtop Pipettor Instruction Manual (CLS-AN-671DOC), visit www.corning.com/lifesciences. Instrument Identification Installation Liquid Head Touch Screen Controller Instrument Deck Disposable Tips and Plates (not included) 1.
  • Page 2: Operating The Unit

    Reagent Reservoir in Position CC ◗ Plates in Position BB and Position AA ◗ After powering up the Corning® Lambda™ EliteMax Benchtop Pipettor, you will be presented with the main screen or Home Screen. Simply touch the desired action to access the submenu.
  • Page 3 Creating User Protocols You can also interact with the plate diagram on the right and modify the rows filled on the destination plate. When you By pressing the Plate Filling, Serial Dilution, or Plate-to-Plate are finished customizing your protocol, return to the General Transfer button in the main menu, the following screen will appear: Settings tab and select Run in the upper right corner.
  • Page 4: Cleaning And Maintenance

    Linked File Information Under the table are settings to edit the table controls: If you would like to save your protocol for future use, check the Auto-keypad: An alphanumeric keypad will automatically pop up ◗ Link File box in the upper right corner. When selected, the Linked when double tapping a table cell.
  • Page 5: Warranty

    Corning’s products are not specifically designed and tested for diagnostic testing. Many Corning products, though not specific for diagnostic testing, can be used in the workflow and preparation of the test at the customers discretion. Customers may use these products to support their claims. We cannot make any claims or statements that our products are approved for diagnostic testing either directly or indirectly.