Table of Contents Warning Safety Information Package Contents Unpacking Installation Intended Use Symbols and Conventions Product Specifications Getting Started 10.0 Important Guidelines 11.0 Operation 12.0 Troubleshooting 13.0 Cleaning and Maintenance 14.0 Temperature Re-adjustment 15.0 Product Disposal for Corning Customers and Recyclers...
WARNING Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user's manual. The equipment is designed only for its intended application. Important safety symbols used in this manual: Warning! This is a warning of risk situations and dangers. Failure to observe this warning could be life-threatening.
If you suspect damage to the unit, immediately file a claim with the carrier in accordance with their instructions before contacting Corning Customer Service. Also verify that all accessories are included and that the unit is in good working order before discarding the shipping packaging.
PRODUCT SPECIFICATIONS Temperature range Ambient +5°C to 99°C Controls PID control Chamber maximum load 20 lbs./9.09 kg Exterior dimensions (W x D x H) 14.9 x 11.4 x 9.0 inches (12.9 inches with cover)/ 38 x 29 x 23 mm (33 mm with cover) Weight 19 lbs./8.6 kg Electrical...
11.0 OPERATION Do not use this water bath in explosive or flammable environments. Do not heat or incubate flammable, explosive, or highly reactive materials in this water bath as serious injury may result. Temperature of water bath cover and chamber edges may reach over 100°C. Use protective gloves to avoid burns and injuries.
12.0 TROUBLESHOOTING Review the information in the table below to troubleshoot operating problems. Problem on Display Cause Solution Temperature probe is not in Examine temperature probe. range Faulty control board. Temperature is more than Examine SET temperature. Examine parameters for heating. 5°C higher than the set Examine parameters for heater.
(plus or minus). After 5 of defining the temperature – display will automatically show temporary temperature. 15.0 PRODUCT DISPOSAL INFORMATION FOR CORNING CUSTOMERS AND RECYCLERS According to Directive 2012/19/EU of the European Parliament and Council of 4...
Page 9
For more specific information on claims, visit the Certificates page at www.corning.com/lifesciences. Warranty/Disclaimer: Unless otherwise specified, all products are for research use only. Not intended for use in diagnostic or therapeutic procedures. Corning Life Sciences makes no claims regarding the performance of these products for clinical or diagnostic applications.
Need help?
Do you have a question about the LSE 6783 and is the answer not in the manual?
Questions and answers