Integra VIAFLO Quick Start Manual page 3

Electronic pipettes

Advertisement

Running a program
Insert the GRIPTIPS into the liquid to be transferred. Press and release RUN
to aspirate the volume selected in the first step (marked with a triangle on the
Run screen).
Your pipette will prompt you to press RUN at each successive step.
Two step blowout
During aspiration, the piston of your electronic pipette moves up (1).
1
During dispensing, the piston returns to the initial position (0). During the last
dispense of a program, it automatically moves further down (2) and blows the
0
remaining liquid from the tip (Blowout).
2
When the piston returns to the initial position (0), a small amount of air is
aspirated, provided the tip is no longer immersed in the liquid (Blowin).
Note: Manually delay the blowin by holding RUN pressed during the last dispense. Remove the
tips from the vessel and release RUN to start blowin.
Tip spacing (VOYAGER only)
In each pipetting mode, the available tip spacing positions are displayed at
the bottom of the screen. The current position is highlighted.
At any program step press  or  to move the tip position to the next spacing.
To modify the tip spacing positions, go to the "Tip Spacing" menu,
accessible from the main menu of the pipette or within every pipetting mode.
Select the positions to be changed (first, middle, last) and press OK. Use 
and  to move the tips to the desired position, click OK and then Save .
Equipment Disposal
Electronic pipettes must not be disposed of with unsorted municipal waste.
Dispose of electronic pipettes in accordance with the laws and regulations in your area
governing disposal of devices.
In certain regions and countries, e.g. in EU member states, the distributor is obliged to
take back this product free of charge at the end of life. Please contact your local
distributor for more details.