Bolus
3.
Using the on-screen keypad enter
desired units of insulin. Tap
⚠ Alert:
Always confirm that the decimal
point placement is correct.
4.
Tap
to confirm the units of insulin
to be delivered.
Note: When administering a Food Bolus,
it is recommended that the user enter a BG
value whenever possible.
Note: If certain advanced features are
being used, the current Dexcom G6 CGM
reading may auto-populate to the bolus
calculator. Please see the user guide for
more information.
5.
Confirm settings.
• Tap
if entered data is correct.
• Tap
to make changes.
6.
Tap
.
The BOLUS INITIATED screen is is
3
temporarily displayed.
After the bolus delivery is complete,
3
an icon displays below the
CGM graph.
Note: If the Control-IQ feature suspends
insulin delivery during a Food Bolus, the remainder of the bolus WILL be delivered before all
insulin delivery is stopped.
90
.
1
4
7
.
INSULIN
10.5
units
Insulin
BG
Units To Deliver
Deliver 10.5 u Bolus?
Correction
EXTENDED
Adjusted Food
25
units
2
5
8
0
10.5
units
GLUCOSE
Add BG
View Calculation
Confirm Request?