Tandem t:slim x2 Training Manual page 84

Insulin pump with control-iq technology
Hide thumbs Also See for t:slim x2:
Table of Contents

Advertisement

Load
12. Fill Tubing is complete. Ensure that
there are no leaks from the tubing
connections. Tap Fill Cannula.
13. Insert a new infusion set and connect
filled tubing to site, then tap
Note: Refer to the infusion set instructions
for proper insertion.
14. Tap Edit Fill Amount if the user wants
to change it. Otherwise, tap START
and skip the next two steps.
15. Select amount needed for cannula fill.
Once a value is selected, the pump
3
will return to the previous screen.
Note: Refer to the infusion set instructions
for proper cannula fill amount.
Note: If the amount needed is not listed,
tap Other amount and use the on-screen
keypad to enter a value between 0.1 to
1.0 units.
82
Change Cartridge
Fill Tubing
Fill Cannula
Site Reminder
Insert a new set and connect to
the filled tubing.
.
On the next screen you will fill the
cannula with insulin.
Edit Fill Amount
Fill Cannula
Amount Filled
0.3 u
0.5 u
0.7 u
Other amount
Load
Press to Turn On
Fill Cannula
0.3 u
START
0
Fill Amount
For Professional Use Only
u