10.0
8.9
6.25
3.9
mmol/L
Control-IQ Technology on the Screen
Control-IQ Technology Status Icon
When Control-IQ technology is on, the CGM Trend Graph displays a diamond icon
in the top left corner. This icon uses different colors to communicate information
about how Control-IQ technology is operating.
When Control-IQ technology is on but not active (i.e. insulin is being delivered
normally), the diamond icon is gray as depicted below. Regardless of the color,
the icon always appears in the same place.
100%
INSULIN ON BOARD
For Professional Use Only
Delivers an automatic correction bolus if sensor glucose
is predicted to be above 10.0 mmol/L
Increases basal insulin delivery if sensor glucose
is predicted to be above 8.9 mmol/L
Maintains active Personal Profile settings
Decreases basal insulin delivery if sensor glucose
is predicted to be below 6.25 mmol/L
Stops basal insulin delivery if sensor glucose
is predicted to be below 3.9 mmol/L
16:45
14 Nov
1.1 u
OPTIONS
BOLUS
Control-IQ Technology
235 u
22
7.0
18
mmol/L
14
10
3
6
HRS
2
111