Fluke 5560A Operator's Manual page 63

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

Test
Steps
(Switch Position)
27
E
28
DC Millivolts
29
30
E
31
Capacitance
32
33
G Diode Test
34
G
[1]
Diode Test
35
F
36
AC Milliamps
37
?
38
AC Amps
39
I
40
m
DC Milliamps
41
42
A
[1]
DC Amps
43
[1] Push the YELLOW button to access this function.
[2] Does not include test lead resistance.
Table 15. Performance Test - 77 Series IV DMM (cont.)
Calibrator Output
30 mV dc
-300 mV dc
600 mV dc
900 nF
9 µF
[1]
90 µF
2.0 V
[1]
0.5 mA 45 Hz
50 mA 1 kHz
400 mA 1 kHz
4.0 A 45 Hz
9.0 A 1 kHz
3 mA, 0 Hz
50 mA
[1]
-400 mA
4.0 A
-9.0 A
Meter Response
Lower Limit
29.8 mV dc
-301.0 mV dc
598.1 mV dc
887 nF
8.87 µF
88.7 µF
1.978 V dc
0.47 mA ac
48.73 mA ac
389.8 mA ac
3.898 A ac
8.75 A ac
2.93 mA dc
49.23 mA dc
-406.2 mA dc
3.938 A dc
-9.16 A dc
Calibrator
Sample Applications
Upper Limit
30.2 mV dc
-299.0 mV dc
601.9 mV dc
913 nF
9.13 µF
91.3 µF
2.022 V dc
0.53 mA ac
51.27 mA ac
410.2 mA ac
4.102 A ac
9.25 A ac
3.07 mA dc
50.77 mA dc
-393.8 mA dc
4.062 A dc
-8.84 A dc
57